EZPass Suture Retriever, Nitinol, Single Pack, Part number 10007380. Orthopedic Manual Surgical Instrument...

Zimmer GmbH / initiated 2014-04-15 / Terminated / root cause: design

Biomet, Inc. began a recall of EZPass Suture Retriever, Nitinol, Single Pack, Part number 10007380. Orthopedic Manual Surgical Instrument, Sterile, single use arthroscopic instrument... on . FDA classified it as Class II and gives the reason as "The nitinol wire or monofilament may buckle within the EZPass Suture Passer and become kinked making the wire/monofilament difficult to use". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1801-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
EZPass Suture Retriever, Nitinol, Single Pack, Part number 10007380. Orthopedic Manual Surgical Instrument, Sterile, single use arthroscopic instrument utilized in conjunction with a nitinol wire to aid in passing suture through soft tissue.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The nitinol wire or monofilament may buckle within the EZPass Suture Passer and become kinked making the wire/monofilament difficult to use.
Root cause tag
design
FDA root cause
Device Design
Action
On 2/28/14, customers received a letter stating these devices have been recalled following an investigation which identified that the nitinol wire or monofilament may buckle within the EzPass Suture Passer and become kinked making the wire/monofilament difficult to use. The letter also stated, there is no adverse health outcome expected for the patient and this action requires the immediate location and discontinued use of the product and its return to Biomet.
Quantity in commerce
100 units
Distribution
Nationwide Distribution.
Product codes and lots
Part numbers 110007380 Lot 223570 and 803860

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1801-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1801-2014Class IITerminatedVoluntary: Firm initiated
Z-1802-2014Class IITerminatedVoluntary: Firm initiated
Z-1803-2014Class IITerminatedVoluntary: Firm initiated