Z-1638-2014Class II

MAMMOMAT Inspiration. Intended for mammography exams, screening, diagnosis, and stereotactic biopsies under...

Siemens Medical Solutions USA, Inc. / initiated 2014-04-25 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of MAMMOMAT Inspiration. Intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals on . FDA classified it as Class II and gives the reason as "There is a potential and possible hazard to the user when using the MAMMOMAT Inspiration PC monitor at the control desk, in that the holder of the PC monitor can break causing an unstable monitor to fall causing...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1638-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
MAMMOMAT Inspiration. Intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals.
Product code
MUE Full Field Digital, System, X-Ray, Mammographic
Reason for recall
There is a potential and possible hazard to the user when using the MAMMOMAT Inspiration PC monitor at the control desk, in that the holder of the PC monitor can break causing an unstable monitor to fall causing possible serious injury.
Root cause tag
design
FDA root cause
Device Design
Action
A safety advisory notice, dated April 25, 2014, was sent to direct accounts to notify them of the issue and provide instructions to avoid the potential risk of injury.
Quantity in commerce
5
Distribution
Distributed in the states of GA, FL, NY, NV, and MA.
Product codes and lots
(Material number 10549894) with serial numbers 5000 through 5029.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1638-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1638-2014Class IITerminatedVoluntary: Firm initiated