510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1948-2014Class I

Manual Resuscitator Bag Series: AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000...

Salter Labs / initiated 2014-05-16 / Terminated

Ventlab LLC began a recall of Manual Resuscitator Bag Series: AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM1000, CPRM2000, CPRM3000, PRO-1900... on . FDA classified it as Class I and gives the reason as "The pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delivered to the victim". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1948-2014
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Salter Labs
Product
Manual Resuscitator Bag Series: AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM1000, CPRM2000, CPRM3000, PRO-1900, PRO-1000, PRO-2000, SC7000, SC8000, SC8020, SC8120, SC9000, SS3200, VN2000, VN3000, VN4000, VN5000, VT1000 , VN2102, VN2002. Emergency manual resuscitator for the Pediatric Population.
Product code
BTM Ventilator, Emergency, Manual (Resuscitator)
Reason for recall
The pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delivered to the victim.
FDA root cause
Process change control
Action
The firm, Ventlab, issued a press dated May 14, 2014, stating Ventlab, LLC has initiated a voluntary medical device removal of a limited number of Ventlab" Resuscitator Bags. The bag series and lot numbers were included in a table as well as manufacture dates. In addition, Ventlab, LLC sent an "URGENT: MEDICAL DEVICE PRODUCT REMOVAL IMMEDIATE ACTION REQUIRED" letter dated May 19, 2014 to its distributors and customers. The letter described the product, problem and actions to be taken. The customers were instructed to review and examine their inventory for affected product; stop using them and immediately contact Ventlab, LLC for further instructions on the return of these products; and if product was further distributed, identify their customers/consignees and notify them of this product removal. The customers were also instructed to complete and return the attached Return Response Form via fax at 1-800-400-8820, Attn: Quality Department, or email to: [email protected] as soon as possible and inform all affected personnel of this removal. Ventlab will send new corrected replacement resuscitation bags to customers once you return the affected product. If you have any questions regarding this action, please call Ventlab at 1-800-237-5481 between the hours of 8:30 AM to 5:00 PM (EST) Monday through Friday, or 1-844-635-5326 or via e-mail at [email protected].
Quantity in commerce
251,384
Distribution
US Nationwide Distribution in the states of OH, NY, TX, TN, FL, CT, and VA,
Product codes and lots
Manual Resuscitator Bag Series: AF1000, AF2000, AF5000, BT4000, VN2000, VN5000, VT1000; Lot #s105147 -107609. AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, PRO-1900, SC7000, SC8120, SS3200, VN2000, VN3000, VN4000, VN5000; Lot #s 106245 -107291. VN2102; Lot #200349. AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM2000, CPRM3000, PRO-1000, PRO-2000, SC7000, SC8020, VN2000, VN3000, VN4000; Lot #s 107029 -107634. VN2002; Lot # 200492. SC8000, SC9000; Lot #s 101441 -107461. CPRM1000; Lot #s 99523 -107315.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1948-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1948-2014Class ITerminatedVoluntary: Firm initiated