Abacus 3.1, Pharmacy Calculator
Baxter Corporation Englewood began a recall of Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding on . FDA classified it as Class II and gives the reason as "Baxter Corporation is initiating a field correction following an upgrade to ABACUS Software v3.1, when Clinimix 4.25110 is used in an order, the Nutritional Summary and any labels containing energy information may...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1847-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Corporation Englewood
- Product
- Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.
- Product code
- NEP System/Device, Pharmacy Compounding
- Reason for recall
- Baxter Corporation is initiating a field correction following an upgrade to ABACUS Software v3.1, when Clinimix 4.25110 is used in an order, the Nutritional Summary and any labels containing energy information may display 1,000 times less protein than is actually present in the solution.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Baxter sent an Urgent Device Correction letter dated May 27, 2014 via certified mail to affected customers. The letter identified the affected product, problem and actios to be taken. Customers were instructed to: 1. Review and distribute this communication to pharmacy staff and compounding staff that use ABACUS software. This notice is intended only for customers that have upgraded to version 3.1 of the software. 2. Please acknowledge receipt of this notification by completing the enclosed customer reply form and return it to Baxter by either faxing it to 1-224-270-5457 or scanning and e-mailing it to [email protected]. 3. If you have distributed ABACUS to other facilities, please notify the end users in accordance with your customary procedures. For questions call Baxter Techenical Support line at 1-800-678-2292.
- Quantity in commerce
- 42
- Distribution
- US Nationwide Distribution and the state of VA.
- Product codes and lots
- Product Code: 8300-0167, 8300-0168, 8300-0169