Z-1847-2014Class II

Abacus 3.1, Pharmacy Calculator

Baxter Corporation Englewood / initiated 2014-05-28 / Terminated / root cause: software

Baxter Corporation Englewood began a recall of Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding on . FDA classified it as Class II and gives the reason as "Baxter Corporation is initiating a field correction following an upgrade to ABACUS Software v3.1, when Clinimix 4.25110 is used in an order, the Nutritional Summary and any labels containing energy information may...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1847-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Corporation Englewood
Product
Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.
Product code
NEP System/Device, Pharmacy Compounding
Reason for recall
Baxter Corporation is initiating a field correction following an upgrade to ABACUS Software v3.1, when Clinimix 4.25110 is used in an order, the Nutritional Summary and any labels containing energy information may display 1,000 times less protein than is actually present in the solution.
Root cause tag
software
FDA root cause
Software design
Action
Baxter sent an Urgent Device Correction letter dated May 27, 2014 via certified mail to affected customers. The letter identified the affected product, problem and actios to be taken. Customers were instructed to: 1. Review and distribute this communication to pharmacy staff and compounding staff that use ABACUS software. This notice is intended only for customers that have upgraded to version 3.1 of the software. 2. Please acknowledge receipt of this notification by completing the enclosed customer reply form and return it to Baxter by either faxing it to 1-224-270-5457 or scanning and e-mailing it to [email protected]. 3. If you have distributed ABACUS to other facilities, please notify the end users in accordance with your customary procedures. For questions call Baxter Techenical Support line at 1-800-678-2292.
Quantity in commerce
42
Distribution
US Nationwide Distribution and the state of VA.
Product codes and lots
Product Code: 8300-0167, 8300-0168, 8300-0169

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1847-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1847-2014Class IITerminatedVoluntary: Firm initiated