The Stryker NAV3i Platform is a mobile cart platform consisting of the sub-components Stryker Nav3 Graphite...
Stryker Instruments Div. of Stryker Corporation began a recall of The Stryker NAV3i Platform is a mobile cart platform consisting of the sub-components Stryker Nav3 Graphite Camera, SPC-3 Computer and IO-Tablet (touch screen)... on . FDA classified it as Class II and gives the reason as "The Stryker NAV3i Platform Power Boxes produced between December 9, 2013 and March 7, 2014 may have been assembled without electrical insulation of the soldering joints". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1715-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- The Stryker NAV3i Platform is a mobile cart platform consisting of the sub-components Stryker Nav3 Graphite Camera, SPC-3 Computer and IO-Tablet (touch screen) as well as the monitor, articulated arms and cart housing. This Navigation system is used in combination with surgical software for cranial, spine, ENT, orthopedic and trauma procedures.
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- The Stryker NAV3i Platform Power Boxes produced between December 9, 2013 and March 7, 2014 may have been assembled without electrical insulation of the soldering joints
- FDA root cause
- Process change control
- Action
- On May 13, 2014, Stryker Instruments sent URGENT MEDICAL DEVICE RECALL notification to their consignees. Notification included the description of the affected product, reason for the recall, risk to health, and actions to be taken by the Customer/User. Once Stryker receives the response from the customer, a Stryker Representative will contact the customer to set up a time to perform the upgrade to the affected power box. Old Power Boxes will be returned to Stryker and placed in an industrial compactor/recycler for destruction. Questions regarding the recall are directed to Kelly Jo Whipple M-F 8am-5pm @ 269-389-2921 or [email protected]
- Quantity in commerce
- 20
- Distribution
- Distributed in the states of AR, CA, IN, LA, MA, MD, MI, MT, NV, and TX.
- Product codes and lots
- Stryker Product Number: 7700-800-000 Stryker Serial Numbers: 15097, 15098, 15099, 15102, 15103, 15104, 15105, 15106, 15107, 15109, 15110, 15111, 15112, 15113, 15114, 15115, 15116, 15117, 15118
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1715-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1716-2014 | Class II | Terminated | Voluntary: Firm initiated |