510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1716-2014Class II

The NAV3i Platform Power Box supplies the electrical components with power (e.g. Stryker Nav3 Graphite...

Stryker Corp. / initiated 2014-05-13 / Terminated

Stryker Instruments Div. of Stryker Corporation began a recall of The NAV3i Platform Power Box supplies the electrical components with power (e.g. Stryker Nav3 Graphite Camera, SPC-3 Computer, IO-Tablet (touch screen) and the... on . FDA classified it as Class II and gives the reason as "The Stryker NAV3i Platform Power Boxes produced between December 9, 2013 and March 7, 2014 may have been assembled without electrical insulation of the soldering joints". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1716-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
The NAV3i Platform Power Box supplies the electrical components with power (e.g. Stryker Nav3 Graphite Camera, SPC-3 Computer, IO-Tablet (touch screen) and the monitor).
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
The Stryker NAV3i Platform Power Boxes produced between December 9, 2013 and March 7, 2014 may have been assembled without electrical insulation of the soldering joints
FDA root cause
Process change control
Action
On May 13, 2014, Stryker Instruments sent URGENT MEDICAL DEVICE RECALL notification to their consignees. Notification included the description of the affected product, reason for the recall, risk to health, and actions to be taken by the Customer/User. Once Stryker receives the response from the customer, a Stryker Representative will contact the customer to set up a time to perform the upgrade to the affected power box. Old Power Boxes will be returned to Stryker and placed in an industrial compactor/recycler for destruction. Questions regarding the recall are directed to Kelly Jo Whipple M-F 8am-5pm @ 269-389-2921 or [email protected]
Quantity in commerce
20
Distribution
Distributed in the states of AR, CA, IN, LA, MA, MD, MI, MT, NV, and TX.
Product codes and lots
Stryker Product Number: 7700-853-000 Stryker Serial Numbers: 10004

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1716-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1715-2014Class IITerminatedVoluntary: Firm initiated
Z-1716-2014Class IITerminatedVoluntary: Firm initiated