Z-1736-2014Class II

VITROS 5,1 FS Chemistry System and VITROS 5,1 FS Refurbished, Product Codes/Catalog Numbers 6801375, 6801890...

Ortho-Clinical Diagnostics, Inc. / initiated 2014-04-10 / Terminated / root cause: software

Ortho-Clinical Diagnostics began a recall of VITROS 5,1 FS Chemistry System and VITROS 5,1 FS Refurbished, Product Codes/Catalog Numbers 6801375, 6801890 for in vitro diagnostic use on . FDA classified it as Class II and gives the reason as "Potential for biased results when a Multiple Window (MW) code is associated with results generated by the VITROS 4600 Chemistry System, the VITROS 5600 Integrated System and the VITROS 5,1 Chemistry System. MW codes may...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1736-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS 5,1 FS Chemistry System and VITROS 5,1 FS Refurbished, Product Codes/Catalog Numbers 6801375, 6801890 for in vitro diagnostic use.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Potential for biased results when a Multiple Window (MW) code is associated with results generated by the VITROS 4600 Chemistry System, the VITROS 5600 Integrated System and the VITROS 5,1 Chemistry System. MW codes may be associated with results from any of the following VITROS Chemistry Products using multiple point (multipoint) rate measurements. The MicroSlide assays that could be affected
Root cause tag
software
FDA root cause
Software design
Action
Important Product Correction Letters (CL2014-116, dated 4/10/2014) were sent on 4/10/2014 via FedEx overnight courier or ORTHO PLUS e-Communications to all customers with VITROS 5,1 FS, 5,1 FS Refurbished, 4600,and/or 5600 Systems. They were informed of the issue and advised to: -- Review VITROS ALT, ALTJ, ALKP, ALKPJ, AST, ASTJ, CHE, CHEJ, CK, CKJ, GGT, GGTJ, LDH and LDHJ results generated on VITROS 4600/5600 and 5,1 FS Systems. Do not report any results associated with an MW code. -- Perform repeat testing in singlicate on all samples with results having an MW code. Only results without an MW code are acceptable. -- Review their current Laboratory Information System (LIS) configuration to determine if changes are required in order to restrict reporting of results with an MW code. Customers with questions should contact Customer Technical Services at 1-800-421-3311. --- Foreign affiliates were informed of the issue by e-mail on 4/10/2014 and instructed to notify their consignees of the issue and actions.
Quantity in commerce
Domestic - 597 units, Foreign - 1381 units
Distribution
Worldwide distribution including US Nationwide, Puerto Rico, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
Product codes and lots
Software Version 2.7 and below; VITROS 5,1 FS Chemistry System Serial Numbers: J34000102 through J34002323; VITROS 5,1 FS Chemistry System Refurbished Serial Numbers: ¿34000337, ¿34001840, ¿34000316, ¿34001389,  34001317, ¿34000538, ¿34000552, ¿34000360, ¿34000451, ¿34000833, ¿ 34001572, ¿34000193, ¿34001633, ¿34000851, ¿34001345, ¿34001396,  34001809, ¿34001897, ¿34001240, ¿34001612, ¿34001311, ¿34000212, 34001472, ¿34001172, ¿34001390, ¿34000557, ¿34000893, ¿34001297, 34001378, ¿34001218, ¿34000318, ¿34000596, ¿34000497, ¿34000495,  34000529, ¿34000478, ¿34001569, ¿34001243, ¿34001504, ¿34000947, 34001422, ¿34000805, ¿34000631, ¿34001248, ¿34001251, ¿34000601, 34001274, ¿34001668, ¿34001850, ¿34001379, ¿34000133, ¿34000642,  34001626.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1736-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1734-2014Class IITerminatedVoluntary: Firm initiated
Z-1735-2014Class IITerminatedVoluntary: Firm initiated
Z-1736-2014Class IITerminatedVoluntary: Firm initiated