Z-2544-2014Class II

Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E...

Medtronic, Inc. / initiated 2014-03-13 / Terminated / root cause: design

Medtronic MiniMed Inc. began a recall of Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722... on . FDA classified it as Class II and gives the reason as "Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2544-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. MiniMed 530G: MMT-551, and MMT-751 Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes.
Product code
OYC Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor
Reason for recall
Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Medtronic, sent an "IMPORTANT MEDICAL DEVICE SAFETY INFORMATION" letter dated March 13, 2014 to its customers. The letter described the product, problem and actions to be taken. The letter informs the customers that because accidental button pressing errors may occur it is important that customers always confirm the insulin dose flashing on the display is correct before pressing ACT to start delivery. The letter informs the customers of what safety settings will minimize the risk of an accidental programming error and what actions to be taken. Should you have any questions about this letter or concerns with your keypad, please do not hesitate to contact the 24 hour helpline at (866) 222-7304 for a replacement.
Quantity in commerce
559,374 units total (444,374 units in US)
Distribution
Worldwide Distribution-Nationwide in US
Product codes and lots
Paradigm: MMT-523NAB, MMT-723KNB, MMT-523NAH, MMT-723KNH, MMT -523NAL, MMT-723KNL, MMT-523NAP, MMT-723KNP, MMT-523NAS, MMT-723KNS, MMT-723NAB, MMT-723NAH, MMT-723NAL, MMT-723NAP, MMT-723NAS, MMT-523KNB, MMT-523KNH, MMT-523KNL, MMT-523KNP, MMT-523KNS.  MiniMed 530G: MMT-551NAM, MMT-551NAH, MMT-551NAL, MMT-551NAP, MMT-551NAS, MMT-751NAB, MMT-751NAH, MMT-751NAL, MMT-751NAP, MMT-751NAS.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2544-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2544-2014Class IITerminatedVoluntary: Firm initiated