Z-1664-2014Class II
Visitec Anterior Chamber Cannula, .40x22mm (27Gx7/8in) REF 585006 Cannulae are intended to irrigate and or...
Becton Dickinson & Co / initiated 2014-05-06 / Terminated / root cause: packaging
Beaver-Visitec International Inc. began a recall of Visitec Anterior Chamber Cannula, .40x22mm (27Gx7/8in) REF 585006 Cannulae are intended to irrigate and or aspirate ophthalmic solutions and liquids in or out... on . FDA classified it as Class II and gives the reason as "Contains a sharp tip needle when the product should contain a blunt tip cannula". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1664-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- Visitec Anterior Chamber Cannula, .40x22mm (27Gx7/8in) REF 585006 Cannulae are intended to irrigate and or aspirate ophthalmic solutions and liquids in or out of the eye, and remove materials where appropriate. These products are used in all sections of the eye. This cannula is a general use cannula mainly designed for the precise maintenance of optimum anterior chamber depth
- Product code
- HMX Cannula, Ophthalmic
- Reason for recall
- Contains a sharp tip needle when the product should contain a blunt tip cannula
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Beaver Visitec (BVI) sent a Product Recall Notice dated May 6, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately return any un-opened product of the affected lot so that BVI can replace the product or issue credit to their company. Customers were asked to fax or email the customer response form to 866-906-4304 or [email protected] as acknowledgement of the letter. Customers with questions should call 781-906-7986. For questions regarding this recall call 781-906-7950.
- Quantity in commerce
- 9970 units
- Distribution
- Nationwide Distribution including AL, CT, IN, NC, NY, TX, and VA.
- Product codes and lots
- Lot number: 3079324
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1664-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY