510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1785-2014Class I

Convatec Flexi Seal Control Fecal Management System Rx Only REF 411107

ConvaTec Limited / initiated 2014-04-29 / Terminated / root cause: design

Convatec Inc. began a recall of Convatec Flexi Seal Control Fecal Management System Rx Only REF 411107 Product Usage:For use to manage fecal incontinence through the collection of liquid to... on . FDA classified it as Class I and gives the reason as "Convatec Inc. is recalling Flexi-seal FMS Control Kit since it does not have a 510K. In addition, an internal assessment of product performance, including a review of customer feedback, has confirmed that Flexi-Seal...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1785-2014
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConvaTec Limited
Product
Convatec Flexi Seal Control Fecal Management System Rx Only REF 411107 Product Usage:For use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications as prescribed by a physician.
Product code
KNT Tubes, Gastrointestinal (And Accessories)
Reason for recall
Convatec Inc. is recalling Flexi-seal FMS Control Kit since it does not have a 510K. In addition, an internal assessment of product performance, including a review of customer feedback, has confirmed that Flexi-Seal Control is not meeting expectations or those of customers. Specifically, the Auto-Valve feature that is unique to the Flexi-Seal Control has not been consistently performing relative
Root cause tag
design
FDA root cause
Device Design
Action
Convatec Inc. sent an Urgent Voluntary Medical Device Recall letter dated April 29, 2014 via UPS. The letter identified the affected product, problem and actions to be taken. Customers are asked to immediately stop distributing and quarantine all recalled lots. Affected products must be returned to Distributor. Contact your Distributor for shipping instruction. Customers should completed the attached return response form and return it to the address on the response form. Questions can be directed to Gregory Kowzun at (336) 547-3710. Convatec issued a press release on 7/3/2014.
Quantity in commerce
73,021 units US; 24,866 units OUS
Distribution
Worldwide Distribution - USA Nationwide
Product codes and lots
13-FM-0301  13-FM-0302  13FM0303  13-FM-0303  13FM0309  13VM509948  13VM509949  13VM510046  13VM512473  13VM513636  13VM513642  13VM516733  13VM520039  13VM521731  13VM521733  13VM524363  13VM524364  13VM527149  13VM527150  13VM528992  13VM528994  13VM531214  14FM0301  14VM531935  14VM532098

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1785-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1785-2014Class ITerminatedVoluntary: Firm initiated