510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1790-2014Class II

Roche MODULAR ANALYTICS EVO Serum Work Area (SWA), Operator's Manual Software Version 08-02; Filter label...

Roche Diagnostics GmbH / initiated 2014-05-21 / Terminated / root cause: labeling and UDI

Roche Diagnostics Operations, Inc. began a recall of Roche MODULAR ANALYTICS EVO Serum Work Area (SWA), Operator's Manual Software Version 08-02; Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320... on . FDA classified it as Class II and gives the reason as "Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1790-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche MODULAR ANALYTICS EVO Serum Work Area (SWA), Operator's Manual Software Version 08-02; Filter label: Filter (N) Mat. No 03149773001, Contents 1, 768-320 Filter (N). Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Roche sent an Urgent Medical Device Correction (UMDC) letter dated May 20, 2014 via UPS Grount (receipt required) to affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: 1. Follow the instructions in the Precautions Regarding Handling the Aspiration Tube Filter section in this UMDC; 2. Discontinue performing ProCell/Clean Cell aspiration tube filter check as this maintenance item will be performed by your Roche FSR as a part of service maintenance; 3. If you have distributed the products listed within this UMDC to other sites or facilities, please provide them with a copy of the UMDC; and 4. Complete the faxback form enclosed with this UMDC and fax it to 1-888-410-6620. For questions contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Quantity in commerce
462
Distribution
US Nationwide Distribution
Product codes and lots
1305-31, 1305-21, 1305-22, 1308-35, 1308-36, 130833, 130834, 1308-44, 1310-20, 1418-12, 1307-33, 1421-11, 1417-20, 1423-16, 143115, 1428-07, 1430-16, 1430-17, 1423-15, 1429-15, 1305-40, 1541-05, 1539-15, 1548-15, 1544-11, 1431-14, 1548-17, 1545-03, 1547-20, 1557-12, 1544-06, 1548-08, 1429-13, 1303-31, 1303-32, 1556-16, 1549-04, 1548-16, 1540-20, 1538-10, 1563-06, 1556-17, 1539-13, 1539-14, 1571-12, 1572-19, 1684-05, 1570-14, 1571-13, 1679-13, 1684-06, 1683-17, 1570-17, 1684-07, 1684-08, 1685-16, 1572-20, 1673-01, 1691-17, 1691-19, 1691-18, 1698-05, 1698-06, 1691-14, 1728-10, 1694-19, 1711-17, 1708-20, 1709-03, 1601-01, 1698-03, 1601-03, 1705-20, 1714-18, 1714-19, 1847-02, 1686-01, 1733-17, 1733-19, 1733-16, 1736-06, 1843-02, 1570-15, 1849-01, 1855-15, 1857-22, 1860-10, 1691-20, 1696-05, 1567-01, 1860-24, 1853-02, 1855-10, 1855-11, 1858-02, 1858-03, 1860-12, 1860-13, 1860-14, 1736-04, 1849-02, 1853-01, 1860-11, 1857-23, 1857-24, 1739-19, 1857-25, 1857-26, 1861-18, 1861-19, 1861-23, 1853-03, 1860-16, 1860-23, 1860-15, 1864-16, 1860-19, 1985-13, 1989-19, 1996-02, 1861-20, 1865-19, 1307-34, 1992-19, 1999-16, 1999-17, 1864-19, 1980-15, 1535-09, 1999-20, 1993-01, 1996-01, 1999-18, 1905-05, 1905-03, 1921-03, 1925-03, 1860-22, 1860-20, 1905-01, 1921-01, 1858-04, 1860-17, 1905-02, 1921-02, 1418-03, 1418-04, 1925-04, 1905-06, 1999-19, 2050-04, 2050-03, 2050-06, 2050-07, 2050-05, 1537-15, 2052-01, 2052-02, 2050-08, 1725-18, 1694-20, 1733-18, 2056-22, 2056-23, 1419-12, 1422-17, 1307-41, 1536-04, 2065-12, 1905-07, 1905-08, 1864-17, 1864-18, 2061-20, 2065-06, 1736-03, 2065-13, 2076-18, 2068-24, 2080-02, 2071-24, 2071-23, 2068-23, 2081-05, 2081-06, 2070-01, 2079-19, 2080-01, 2081-04, 2070-02, 2086-14, 2086-20, 2084-16, 2084-17, 2084-18, 2085-09, 2086-17, 2086-18, 2082-39, 2084-05, 2084-12, 2084-13, 2084-14, 2084-15, 2085-05, 2085-07, 2085-08, 2188-05, 2188-06, 2188-07, 2086-15, 2086-16, 2086-19, 2188-08, 2188-09, 2188-10, 2084-06, 2084-07, 2071-22, 2084-08, 2084-09, 2084-10, 2084-11, 2085-04, 2085-10, 2188-04, 2084-19, 2084-20, 2084-21, 2076-19, 2076-20, 2190-10, 2190-19, 2190-20, 2190-06, 2190-07, 2190-08, 2190-09, 2190-11, 2190-12, 2081-07, 2081-08, 2081-09, 2081-10, 2081-11, 2083-04, 2083-05, 2084-22, 2084-23, 2076-21, 2190-13, 2083-06, 2083-07, 2083-08, 2079-18, 2081-12, 2190-14, 2080-03, 2188-36, 2188-37, 2188-38, 2188-31, 2188-34, 2188-35, 2187-33, 2188-39, 2188-40, 2190-17, 2190-18, 2190-21, 2082-40, 2083-01, 2188-32, 2190-03, 2190-04, 2190-05, 2190-15, 2190-16, 1549-02, 2065-04, 2071-20, 2071-21, 2189-40, 2190-01, 2190-02, 2083-03, 2085-18, 2189-37, 2189-38, 2189-39, 2085-16, 2189-36, 2192-24, 1561-12, 1544-10, 2083-02, 2085-17, 1537-17, 2187-36, 2187-37, 2187-38, 2102-06, 2102-07, 2102-08, 2085-12, 2192-21, 2192-22, 2086-12, 2106-28, 2106-29, 2110-09, 2110-10, 2189-01, 2085-13, 2192-28, 2192-29, 2085-14, 2192-26, 2192-27, 2192-31, 2086-23, 2086-24, 2102-03, 2195-23, 2085-11, 2085-15, 2192-25, 2192-32, 1711-16, 2086-25, 2086-26, 2101-39, 2188-03, 2189-04, 2188-01, 2188-02, 2119-13, 2189-05, 2189-06, 2086-13, 2086-21, 2189-02, 2189-03, 2102-04, 2102-05, 2119-14, 2192-18, 2192-19, 2126-07, 2132-14, 2192-20, 2126-08, 2189-07, 2140-02, 1545-05, 1686-03, 2146-17, 2146-18, 2146-19, 2251-16, 2251-17, 2251-18, 2252-02, 2252-04, 2253-15, 2147-05, 2147-06, 2251-19, 2251-20, 2252-03, 2253-16, 2253-17, 2253-18, 2260-11, 2260-12, 2260-13, 2257-08, 2257-09, 2257-10, 2257-11, 2252-05, 2253-19, 2253-20, 2254-01, 2255-20, 2257-04, 2056-21, 2147-04, 2251-15, 2253-12, 2253-13, 2253-14, 2257-05, 2257-06, 2257-07, 1725-19, 1857-27, 2076-22, 2260-17, 2260-18, 2260-19, 2260-20, 2262-11, 2262-12, 1725-20, 1691-15, 2260-14, 2260-15, 2260-16, 2256-01, 1684-02, 2256-02, 2262-16, 2262-17, 2263-12, 2262-19, 2262-20, 2267-20, 2263-13, 2263-14, 1847-01, 2071-25, 2075-20, 2262-18, 2275-11, 2275-12, 2267-17, 2267-18, 2263-16, 2263-17, 2275-14, 2290-05, 2267-19, 2275-09, 2275-10, 2275-13, 1556-12, 2275-15, 2275-16, 2278-07, 2290-06, 2278-08, 2283-16, 2296-20, 1563-04, 2278-09, 2263-15, 2278-10, 2278-11, 2283-17, 1534-12, 1601-02, 2330-19, 2331-01, 1544-13, 2330-20, 2331-02, 2330-16, 2330-17, 2330-18, 1559-13, 1307-35.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1790-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1790-2014Class IITerminatedVoluntary: Firm initiated
Z-1791-2014Class IITerminatedVoluntary: Firm initiated
Z-1792-2014Class IITerminatedVoluntary: Firm initiated