Medtronic Model 8780 Ascenda Intrathecal Catheter. The implantable Medtronic Model 8780 Ascenda Intrathecal...
Medtronic Neuromodulation began a recall of Medtronic Model 8780 Ascenda Intrathecal Catheter. The implantable Medtronic Model 8780 Ascenda Intrathecal Catheter is part of an infusion system that stores... on . FDA classified it as Class II and gives the reason as "Medtronic has determined that an Ascenda catheter kit may have had endotoxin in excess of USP limits". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1723-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Neuromodulation
- Product
- Medtronic Model 8780 Ascenda Intrathecal Catheter. The implantable Medtronic Model 8780 Ascenda Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and a Model 8780 Ascenda Intrathecal Catheter.
- Product code
- LKK Pump, Infusion, Implanted, Programmable
- Reason for recall
- Medtronic has determined that an Ascenda catheter kit may have had endotoxin in excess of USP limits.
- Root cause tag
- sterility
- FDA root cause
- Release of Material/Component prior to receiving test results
- Action
- There was only one consignee affected by the recall. The consignee was contacted by phone on 5/2/14 and the information was provided verbally. General Manager & Vice President, Pain Stimulation/Targeted Drug Delivery spoke via phone with the physician managing the patient. The following information was provided: "Medtronic identified the potential for endotoxin levels exceeding USP requirements on product recently implanted (Model 8780, serial number N457062003). "There may be the potential for serious patient injury due to endotoxin within the first hours after implant. "Medtronic recommends observing the patient for symptoms of meningitis. Also to be aware that CSF would not test with a positive culture because endotoxin induced meningitis results in an aseptic meningitis. The physician expressed confidence in the management of the patient and stated he was calling the nursing care facility to increase monitoring of the patient. Subsequent information was received regarding the patient indicating there were no signs of an endotoxin reaction.
- Quantity in commerce
- 1
- Distribution
- Distributed in the state of CT.
- Product codes and lots
- N457062003
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1723-2014 | Class II | Terminated | Voluntary: Firm initiated |