Z-1723-2014Class II

Medtronic Model 8780 Ascenda Intrathecal Catheter. The implantable Medtronic Model 8780 Ascenda Intrathecal...

Medtronic Neuromodulation / initiated 2014-05-02 / Terminated / root cause: sterility

Medtronic Neuromodulation began a recall of Medtronic Model 8780 Ascenda Intrathecal Catheter. The implantable Medtronic Model 8780 Ascenda Intrathecal Catheter is part of an infusion system that stores... on . FDA classified it as Class II and gives the reason as "Medtronic has determined that an Ascenda catheter kit may have had endotoxin in excess of USP limits". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-1723-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
Medtronic Model 8780 Ascenda Intrathecal Catheter. The implantable Medtronic Model 8780 Ascenda Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and a Model 8780 Ascenda Intrathecal Catheter.
Product code
LKK Pump, Infusion, Implanted, Programmable
Reason for recall
Medtronic has determined that an Ascenda catheter kit may have had endotoxin in excess of USP limits.
Root cause tag
sterility
FDA root cause
Release of Material/Component prior to receiving test results
Action
There was only one consignee affected by the recall. The consignee was contacted by phone on 5/2/14 and the information was provided verbally. General Manager & Vice President, Pain Stimulation/Targeted Drug Delivery spoke via phone with the physician managing the patient. The following information was provided: "Medtronic identified the potential for endotoxin levels exceeding USP requirements on product recently implanted (Model 8780, serial number N457062003). "There may be the potential for serious patient injury due to endotoxin within the first hours after implant. "Medtronic recommends observing the patient for symptoms of meningitis. Also to be aware that CSF would not test with a positive culture because endotoxin induced meningitis results in an aseptic meningitis. The physician expressed confidence in the management of the patient and stated he was calling the nursing care facility to increase monitoring of the patient. Subsequent information was received regarding the patient indicating there were no signs of an endotoxin reaction.
Quantity in commerce
1
Distribution
Distributed in the state of CT.
Product codes and lots
N457062003

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1723-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1723-2014Class IITerminatedVoluntary: Firm initiated