Z-1700-2014Class II

Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The...

Hologic, Inc. / initiated 2014-05-13 / Terminated

Hologic, Inc. began a recall of Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates... on . FDA classified it as Class II and gives the reason as "Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1700-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hologic, Inc.
Product
Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer.
Product code
MUE Full Field Digital, System, X-Ray, Mammographic
Reason for recall
Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury
FDA root cause
Component change control
Action
Hologic notified customers by letter dated 5/13/14 via certified mail or Federal Express to inform customers of the problem and provide acknowledgement of the notification and confirmation that their systems have been corrected A Dealer notification dated 5/13/14 sent by certified mail or Federal Express to inform Dealers of the need to notify their customers of the problem and to correct the identified Selenia system using Hologic Technical Bulletin TB-00806. On 5/16/14 the firm expanded the recall to an additional units Mammography System M-IV (2 units) and Selenia Mammography Systems (4) by letter dated 5/14/14. If there are any questions regarding this field correction, please call Hologic Customer Support at 877 371-4372.
Quantity in commerce
11 units
Distribution
Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.
Product codes and lots
Serial Numbers: 2840314W8065, 2840314W8062W, 2840314W8066W, 2840314W8067 2840314W8073, 2840314W8064, 2840314W8056W, 28401082720 28407083579, 28401137611W, 28403083045W

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1700-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1700-2014Class IITerminatedVoluntary: Firm initiated
Z-1701-2014Class IITerminatedVoluntary: Firm initiated