Ultrasonic Transducer Kits For use with the Access Family of Immunoassay Systems*. *The Access Family of...
Beckman Coulter Inc. began a recall of Ultrasonic Transducer Kits For use with the Access Family of Immunoassay Systems*. *The Access Family of Immunoassay Systems includes the Access 2, UniCel DxI... on . FDA classified it as Class II and gives the reason as "Beckman Coulter is initiating a field action because some Ultrasonic Transducer Kits were potentially assembled with incorrect mounting screws". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1737-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Ultrasonic Transducer Kits For use with the Access Family of Immunoassay Systems*. *The Access Family of Immunoassay Systems includes the Access 2, UniCel DxI 800 and UniCel DxI 600, UniCel DxC 600i, and the UniCel DxC 880i, UniCel DxC 860i, UniCel DxC 680i, and UniCel DxC 660i systems.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Beckman Coulter is initiating a field action because some Ultrasonic Transducer Kits were potentially assembled with incorrect mounting screws.
- FDA root cause
- Mixed-up of materials/components
- Action
- Beckman Coulter sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 8, 2014 via first class mail. The letter identified the reason for recall, product information, issue, impact, action, resolution, contact information, and customer response form. Customers are advised to review historical quality control (QC) data and System Check results, and determine whether a review of patient test results is warranted. For questions regarding this letter, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com/customersupport/support, By phone: call 1-800-854-3633 in the United States and Canada, and Outside the United States and Canada, contact your local Beckman Coulter representative
- Quantity in commerce
- 82 units
- Distribution
- Worldwide Distribution and USA Nationwide - and the countries of Australia, Canada, China, France, Hong Kong, Italy, Japan, Mexico, Portugal, Spain, Switzerland, Taiwan,and Turkey.
- Product codes and lots
- Access 2 Immunoassay Systems, Catalog Number: 81600N. UniCel DxI 600 Access Immunoassay Systems, Catalog Numbers: A30260 and A71460. UniCel DxI 800 Access Immunoassay Systems, Catalog Numbers: 973100 and A71456. UniCel DxC 600i SYNCHRON Access Clinical Systems, Catalog Number: A25640. Catalog numbers for these systems are included in the respective DxI 600 and DxI 800 model numbers: UniCel DxC 660i SYNCHRON Access Clinical Systems, UniCel DxC 680i SYNCHRON Access Clinical Systems, UniCel DxC 860i SYNCHRON Access Clinical Systems, UniCel DxC 880i SYNCHRON Access Clinical Systems.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1737-2014 | Class II | Terminated | Voluntary: Firm initiated |