Z-2717-2014Class II
CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50
KCI USA, Inc. / initiated 2014-05-19 / Terminated / root cause: design
KCI USA, Inc. began a recall of CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50 on . FDA classified it as Class II and gives the reason as "Blade can shift during shipping which makes the product ineffective". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2717-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- KCI USA, Inc.
- Product
- CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50
- Product code
- GFD Dermatome
- Reason for recall
- Blade can shift during shipping which makes the product ineffective.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- KCI sent an Urgent - Voluntary Medical Device Field Notification letter dated June 6, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is sending Field Notifications to affected customers instructing them to complete appropriate actions. Option I: Contact your local KCI Representative or KCI Customer Technical Support at 1-800-275-4524 and select Option 3 to arrange for return and replacement of the affected harvester(s). Option 2: If you prefer, you may re-set the blade after removal from the packaging by gently tapping the curved end of the harvester against the palm of your hand, as shown in Figure 3, until the blade is no longer visible in the harvester top plate, as shown in Figure 4.
- Quantity in commerce
- 1,331 units
- Distribution
- US nationwide Distribution
- Product codes and lots
- Part number CT-H25 with the following lot numbers: 3244680000, 3446210000, 3246280000, 3246480000, 3246490000, 3246970000, 3247020000, 3247030000, 3247130000, 3247440000, 3247670000, 3247680000, 3247980000, 3248070000, 3248200000, 3248300000, 3249120000, 3249130000, 3249140000, 3249150000, 3249160000. Part number CT-H50 with the following lot numbers: 324782, 324796, 324797, 3245120000, 3245670000, 3245680000, 3246270000, 3246490000, 3246690000, 3246700000, 3246940000, 3246980000, 3247140000, 3247160000, 3247170000, 3247250000, 3247310000, 3247470000, 324780000, 3247810000, 3247820000, 3247960000, 3247970000, 3248080000, 3248190000, 3248230000, 3248300000, 3248380000, 3248960000, 3248970000, 3248980000, 3248990000, 3249000000, 3249010000, 3249020000, 3249030000, 3249040000, 3249050000, 3249060000, 3249070000, 3249080000, 3249090000, 3249100000, 3249110000, 3249390000, 3249580000.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2717-2014 | Class II | Terminated | Voluntary: Firm initiated |