Z-2717-2014Class II

CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50

KCI USA, Inc. / initiated 2014-05-19 / Terminated / root cause: design

KCI USA, Inc. began a recall of CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50 on . FDA classified it as Class II and gives the reason as "Blade can shift during shipping which makes the product ineffective". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2717-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
KCI USA, Inc.
Product
CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50
Product code
GFD Dermatome
Reason for recall
Blade can shift during shipping which makes the product ineffective.
Root cause tag
design
FDA root cause
Device Design
Action
KCI sent an Urgent - Voluntary Medical Device Field Notification letter dated June 6, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is sending Field Notifications to affected customers instructing them to complete appropriate actions. Option I: Contact your local KCI Representative or KCI Customer Technical Support at 1-800-275-4524 and select Option 3 to arrange for return and replacement of the affected harvester(s). Option 2: If you prefer, you may re-set the blade after removal from the packaging by gently tapping the curved end of the harvester against the palm of your hand, as shown in Figure 3, until the blade is no longer visible in the harvester top plate, as shown in Figure 4.
Quantity in commerce
1,331 units
Distribution
US nationwide Distribution
Product codes and lots
Part number CT-H25 with the following lot numbers: 3244680000, 3446210000, 3246280000, 3246480000, 3246490000, 3246970000, 3247020000, 3247030000, 3247130000, 3247440000, 3247670000, 3247680000, 3247980000, 3248070000, 3248200000, 3248300000, 3249120000, 3249130000, 3249140000, 3249150000, 3249160000.  Part number CT-H50 with the following lot numbers: 324782, 324796, 324797, 3245120000, 3245670000, 3245680000, 3246270000, 3246490000, 3246690000, 3246700000, 3246940000, 3246980000, 3247140000, 3247160000, 3247170000, 3247250000, 3247310000, 3247470000, 324780000, 3247810000, 3247820000, 3247960000, 3247970000, 3248080000, 3248190000, 3248230000, 3248300000, 3248380000, 3248960000, 3248970000, 3248980000, 3248990000, 3249000000, 3249010000, 3249020000, 3249030000, 3249040000, 3249050000, 3249060000, 3249070000, 3249080000, 3249090000, 3249100000, 3249110000, 3249390000, 3249580000.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2717-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2717-2014Class IITerminatedVoluntary: Firm initiated