Alcon CUSTOM-PAK containing Anterior Chamber Cannula 27G Component Part Number 585006 contained in Alcon...
Alcon Research, Ltd. began a recall of Alcon CUSTOM-PAK containing Anterior Chamber Cannula 27G Component Part Number 585006 contained in Alcon Assembly Numbers 9795-37, 11233-14, and 13334-04. The... on . FDA classified it as Class II and gives the reason as "Alcon CUSTOM-PAK which is supposed to contain a 27G Anterior Chamber Cannula (27G x 7/8 in) actually contains a 27G Sharp Tip Needle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1947-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alcon Laboratories
- Product
- Alcon CUSTOM-PAK containing Anterior Chamber Cannula 27G Component Part Number 585006 contained in Alcon Assembly Numbers 9795-37, 11233-14, and 13334-04. The product is typically inserted through an incision during cataract surgery to deliver balanced salt solution to help anterior chamber space and replace fluids that are lost during surgery.
- Product code
- LRO General Surgery Tray
- Reason for recall
- Alcon CUSTOM-PAK which is supposed to contain a 27G Anterior Chamber Cannula (27G x 7/8 in) actually contains a 27G Sharp Tip Needle
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Alcon sent an Urgert Market Withdrawal letter dated May 14, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory to determine if they have any affected product, place the affected product into quarantine to ensure it is not used, return the attached Response Form via fax to 817-302-4337 or email [email protected]. Customers with questions were instructed to call 1-800-862-5266.
- Quantity in commerce
- 684 total (525 units in US)
- Distribution
- Nationwide Distribution including North Carolina and Maine
- Product codes and lots
- Alcon Assembly Numbers 9795-37, 11233-14, and 13334-04 bearing Lot numbers 1584114H, 1577346H, 1595079H, 1589608H, 1595079H, 1585992H, 1580988H, 1585992H, and 1580492H.