VerSys Hip System, Beaded Fullcoat Stems, Zimmer, Inc., Warsaw, IN 46581
Zimmer, Inc. began a recall of VerSys Hip System, Beaded Fullcoat Stems, Zimmer, Inc., Warsaw, IN 46581 on . FDA classified it as Class II and gives the reason as "During routine inspection, beaded hip stem forgings received from supplier failed to meet ZES 2A-102. Processing of Beaded CoCrMo Alloy Hip Stem, material requirements per ATS #14-04818 and ATS #14-05549, respectively". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1842-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- VerSys Hip System, Beaded Fullcoat Stems, Zimmer, Inc., Warsaw, IN 46581
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- During routine inspection, beaded hip stem forgings received from supplier failed to meet ZES 2A-102. Processing of Beaded CoCrMo Alloy Hip Stem, material requirements per ATS #14-04818 and ATS #14-05549, respectively.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 6/2/2014 the firm sent "Urgent Medical Device Recall" letters to their customers. On 8/13/14, Zimmer contacted a distributor via phone pertaining to one additional lot (00-7843-011-06; lot 62547211) which had been included in the recall. Formal recall notices were issued to the distributor via email on 8/15/14.
- Quantity in commerce
- 147 units
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, MA, MI, MN, NC, ND, NJ, NY, OH, OK, PA, TN, TX, VA & WI, and the countries of Canada, Colombia, Dominican Republic, Japan, Korea (S. Korea), Peru, Taiwan & Thailand.
- Product codes and lots
- Item # 784301108, Lot #'s: 62478967, 62479742, 62503408, 62528282, 62530786, 62544915, 62554360; Item # 784301126, Lot #: 62544885, 62547211, Item # 784301156, Lot #, 62587079; Item # 784301208, Lot #'s: 62503411, 62509385, 62530790, 62547217, 62554362; Item # 784301226, Lot #'s: 62515502, 62547216; Item # 784301256, Lot #'s: 62489572, 62522386; Item # 784301326, Lot #, 62496464, 62594428, 62622893; Item # 784301356, Lot #: 62622894; Item # 784301382, Lot #: 62544907; Item # 784301511, Lot #: 62472423 & Item # 784301581, Lot #: 62496469.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1842-2014 | Class II | Terminated | Voluntary: Firm initiated |