Z-1842-2014Class II

VerSys Hip System, Beaded Fullcoat Stems, Zimmer, Inc., Warsaw, IN 46581

Zimmer GmbH / initiated 2014-05-19 / Terminated / root cause: manufacturing process

Zimmer, Inc. began a recall of VerSys Hip System, Beaded Fullcoat Stems, Zimmer, Inc., Warsaw, IN 46581 on . FDA classified it as Class II and gives the reason as "During routine inspection, beaded hip stem forgings received from supplier failed to meet ZES 2A-102. Processing of Beaded CoCrMo Alloy Hip Stem, material requirements per ATS #14-04818 and ATS #14-05549, respectively". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1842-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
VerSys Hip System, Beaded Fullcoat Stems, Zimmer, Inc., Warsaw, IN 46581
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
During routine inspection, beaded hip stem forgings received from supplier failed to meet ZES 2A-102. Processing of Beaded CoCrMo Alloy Hip Stem, material requirements per ATS #14-04818 and ATS #14-05549, respectively.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 6/2/2014 the firm sent "Urgent Medical Device Recall" letters to their customers. On 8/13/14, Zimmer contacted a distributor via phone pertaining to one additional lot (00-7843-011-06; lot 62547211) which had been included in the recall. Formal recall notices were issued to the distributor via email on 8/15/14.
Quantity in commerce
147 units
Distribution
Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, MA, MI, MN, NC, ND, NJ, NY, OH, OK, PA, TN, TX, VA & WI, and the countries of Canada, Colombia, Dominican Republic, Japan, Korea (S. Korea), Peru, Taiwan & Thailand.
Product codes and lots
Item # 784301108, Lot #'s: 62478967, 62479742, 62503408, 62528282, 62530786, 62544915, 62554360; Item # 784301126, Lot #: 62544885, 62547211, Item # 784301156, Lot #, 62587079; Item # 784301208, Lot #'s: 62503411, 62509385, 62530790, 62547217, 62554362; Item # 784301226, Lot #'s: 62515502, 62547216; Item # 784301256, Lot #'s: 62489572, 62522386; Item # 784301326, Lot #, 62496464, 62594428, 62622893; Item # 784301356, Lot #: 62622894; Item # 784301382, Lot #: 62544907; Item # 784301511, Lot #: 62472423 & Item # 784301581, Lot #: 62496469.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1842-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1842-2014Class IITerminatedVoluntary: Firm initiated