Z-2086-2014Class II
NOVATION Femoral Stem, Tapered Extended offset, plasma- coated, press fit. Catalog number 160-01-13 intended...
Exactech, Inc. / initiated 2014-03-07 / Terminated / root cause: design
Exactech, Inc. began a recall of NOVATION Femoral Stem, Tapered Extended offset, plasma- coated, press fit. Catalog number 160-01-13 intended to be implanted to replace a hip joint with our... on . FDA classified it as Class II and gives the reason as "The femoral stem is lacking specified plasma coating". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2086-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Exactech, Inc.
- Product
- NOVATION Femoral Stem, Tapered Extended offset, plasma- coated, press fit. Catalog number 160-01-13 intended to be implanted to replace a hip joint with our without bone cement.
- Product code
- LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
- Reason for recall
- The femoral stem is lacking specified plasma coating.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Exactech sent an Important Product Recall Notice dated March 10, 2014, to all affected customers. The notice identified the product the problem and the action to be taken by the customer. Description of Issue: The femoral stem is lacking specified plasma spray coating. Clinical Impact: Without plasma spray coating, the stem may not remain stable and will not have the possibility for biological fixation. In order to comply with applicable regulations and Exactech policies customers were instructed to : " Immediately cease distribution or use of these products. " Extend this information to your accounts that may have this product in their possession. " Determine if they have any of the affected devices " Fax back the attached form. Complete the attached fax back form and return it to Exactech. Customers were instructed to contact their Exactech inventory representative within 5 business days to confirm quantities at their location. For any inventory restocking questions related to this issue, please call 1-800-392-2832. For questions regarding this recall call 800-392-2832.
- Quantity in commerce
- 10
- Distribution
- Worldwide Distribution including Spain, Greece and Japan
- Product codes and lots
- Serial Numbers: 2910243, 2910244, 2910245, 2910246, 2910247, 2910248, 2910249, 2910250, 2910251, and 2910252
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2086-2014 | Class II | Terminated | Voluntary: Firm initiated |