510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2086-2014Class II

NOVATION Femoral Stem, Tapered Extended offset, plasma- coated, press fit. Catalog number 160-01-13 intended...

Exactech, Inc. / initiated 2014-03-07 / Terminated / root cause: design

Exactech, Inc. began a recall of NOVATION Femoral Stem, Tapered Extended offset, plasma- coated, press fit. Catalog number 160-01-13 intended to be implanted to replace a hip joint with our... on . FDA classified it as Class II and gives the reason as "The femoral stem is lacking specified plasma coating". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2086-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Exactech, Inc.
Product
NOVATION Femoral Stem, Tapered Extended offset, plasma- coated, press fit. Catalog number 160-01-13 intended to be implanted to replace a hip joint with our without bone cement.
Product code
LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Reason for recall
The femoral stem is lacking specified plasma coating.
Root cause tag
design
FDA root cause
Component design/selection
Action
Exactech sent an Important Product Recall Notice dated March 10, 2014, to all affected customers. The notice identified the product the problem and the action to be taken by the customer. Description of Issue: The femoral stem is lacking specified plasma spray coating. Clinical Impact: Without plasma spray coating, the stem may not remain stable and will not have the possibility for biological fixation. In order to comply with applicable regulations and Exactech policies customers were instructed to : " Immediately cease distribution or use of these products. " Extend this information to your accounts that may have this product in their possession. " Determine if they have any of the affected devices " Fax back the attached form. Complete the attached fax back form and return it to Exactech. Customers were instructed to contact their Exactech inventory representative within 5 business days to confirm quantities at their location. For any inventory restocking questions related to this issue, please call 1-800-392-2832. For questions regarding this recall call 800-392-2832.
Quantity in commerce
10
Distribution
Worldwide Distribution including Spain, Greece and Japan
Product codes and lots
Serial Numbers: 2910243, 2910244, 2910245, 2910246, 2910247, 2910248, 2910249, 2910250, 2910251, and 2910252

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2086-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2086-2014Class IITerminatedVoluntary: Firm initiated