510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1732-2014Class II

RUSCH Easy Cath Coude Kit, 10 FR, Rx Only, Sterile, Distributed by: Teleflex Medical, Research Triangle Park...

Teleflex Medical / initiated 2014-05-21 / Terminated / root cause: packaging

Teleflex Medical began a recall of RUSCH Easy Cath Coude Kit, 10 FR, Rx Only, Sterile, Distributed by: Teleflex Medical, Research Triangle Park, NC 27708. A tubular device that is inserted... on . FDA classified it as Class II and gives the reason as "Product was packaged with a straight catheter rather than the correct coude (curved) catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1732-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Teleflex Medical
Product
RUSCH Easy Cath Coude Kit, 10 FR, Rx Only, Sterile, Distributed by: Teleflex Medical, Research Triangle Park, NC 27708. A tubular device that is inserted through the urethera and is used to pass fluids to or from urinary tracts.
Product code
KOD Catheter, Urological
Reason for recall
Product was packaged with a straight catheter rather than the correct coude (curved) catheter.
Root cause tag
packaging
FDA root cause
Mixed-up of materials/components
Action
Teleflex Medical sent an Urgent Medical Device Recall Notification dated May 21, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to immediately discontinue use and quarantine any of the affected product in stock. They were to complete and return the enclosed Recall Acknowledgement Form via fax to Teleflex Medical. Upon receipt, they will be contacted by a Customer Service Representative with instructions for the return of the product to Teleflex Medical. Teleflex Medical is committed to providing high quality, safe and effective products. We sincerely apologize for any inconvenience this action may cause your operations. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990.
Quantity in commerce
1000 eaches
Distribution
US Distribution including the states of CA, FL, OH and OK.
Product codes and lots
Product Code: ECK103, Lot number: 02A1402484

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1732-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1732-2014Class IITerminatedVoluntary: Firm initiated