Z-1705-2014Class II

Spacelabs Medical Ultraview SL Command Module, Model 91496. Arrhythmia Detector and Alarm. The...

GE Healthcare / initiated 2014-05-08 / Terminated / root cause: software

Spacelabs Healthcare Inc began a recall of Spacelabs Medical Ultraview SL Command Module, Model 91496. Arrhythmia Detector and Alarm. The Multi-parameter Module is used with the Patient Care Management... on . FDA classified it as Class II and gives the reason as "A software anomaly causes temperature alarm limits to change to limits which are not physiologically possible and result in nuisance false positive alarms". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1705-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Spacelabs Medical Ultraview SL Command Module, Model 91496. Arrhythmia Detector and Alarm. The Multi-parameter Module is used with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from a patient in a clinical environment.
Product code
DSI Detector And Alarm, Arrhythmia
Reason for recall
A software anomaly causes temperature alarm limits to change to limits which are not physiologically possible and result in nuisance false positive alarms.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
Spacelabs Healthcare identified a software anomaly in the Ultraview SL" Command Module, Model 91496. During an internal audit, Spacelabs identified a reportable recall that was not reported. We have a product called the Spacelabs Medical Ultraview SL" Command Module, Model 91496. A software error was discovered at a customer site which was not a safety issue, but in the interest of customer service it was decided to downgrade all their modules with an older software version.
Quantity in commerce
64 devices
Distribution
US Distribution to GA.
Product codes and lots
All devices of Model: 91496.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1705-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1705-2014Class IITerminatedVoluntary: Firm initiated