Z-1718-2014Class II

Patient Side Cart assembly, 4-arm, for use with Intuitive Surgical, Inc. da Vinci Surgical System IS 4000...

Intuitive Surgical, Inc. / initiated 2014-05-09 / Terminated / root cause: manufacturing process

Intuitive Surgical, Inc. began a recall of Patient Side Cart assembly, 4-arm, for use with Intuitive Surgical, Inc. da Vinci Surgical System IS 4000. The patient-side cart is positioned at the operating... on . FDA classified it as Class II and gives the reason as "Torque wrenches used to tighten certain PSC bolts were out of calibration. This could lead to over- or under-tightening of bolts that may result in system faults, mechanical sound (e.g. scraping, squeaking, rattling)...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1718-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intuitive Surgical, Inc.
Product
Patient Side Cart assembly, 4-arm, for use with Intuitive Surgical, Inc. da Vinci Surgical System IS 4000. The patient-side cart is positioned at the operating room table and contains four arms that are positioned over the target patient anatomy. The endoscope attaches onto any arm. The patient-side assistant attaches/detaches the endoscope and instrument intra-operatively.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
Torque wrenches used to tighten certain PSC bolts were out of calibration. This could lead to over- or under-tightening of bolts that may result in system faults, mechanical sound (e.g. scraping, squeaking, rattling), cracks or vibration in the system.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Intuitive sent an Urgent Medical Device Correction letter dated May 13, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. All personnel who use the device should be notified of the issue and the acknowledgement form completed and returned. Intuitive reps will visit each affect sites to install a replacement PSC and ensure that it is properly integrated with the Da Vinci. For questions contact Intuitive Surgical at 800-876-1310 Option 3 (Monday - Friday 6 am to 5 pm PST).
Quantity in commerce
2
Distribution
Medical facilities in TX and FL
Product codes and lots
lot 380652-20

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1718-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1718-2014Class IITerminatedVoluntary: Firm initiated