Z-1985-2014Class II

Siemens Mobilett XP Digital Mobile X-ray System

Siemens Medical Solutions USA, Inc. / initiated 2014-05-09 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards... on . FDA classified it as Class II and gives the reason as "There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1985-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards, intensive care, and premature birth-wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.
Product code
IZL System, X-Ray, Mobile
Reason for recall
There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.
Root cause tag
design
FDA root cause
Component design/selection
Action
Siemens sent a Safety Advisory Notice letter dated May 9, 2014, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that a Siemens service technician would be visiting their site to identify and service affected devices.
Quantity in commerce
10
Distribution
US Nationwide Distributions in the states of MO, GA, FL, NJ, CA, PA and including NM.
Product codes and lots
model no. 1818447, with serial nos.: 3284 3449 1026 1025 1010 1041 1054 2002 2058 2022

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1985-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1985-2014Class IITerminatedVoluntary: Firm initiated