MOSAIQ MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery...
Elekta, Inc. began a recall of MOSAIQ MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare... on . FDA classified it as Class II and gives the reason as "MOSAIQ does not re-calculate the agent volume when the user changes the ordering dose (to other than what was calculated) prior to saving the order". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1869-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- MOSAIQ MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used where ever radiotherapy and/or chemotherapy are prescribed.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- MOSAIQ does not re-calculate the agent volume when the user changes the ordering dose (to other than what was calculated) prior to saving the order.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Elekta issued an Important Safety Notice dated May 29, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Corrective action that shall be taken: Notify the affected customers of the issue - To be completed 06/22/2014 Follow up with the affected customers for: Confirmation receipt of the LIN, AND Confirmation of installation of patch for the fix- To be completed 11/30/2014 The notice includes a confirmation which is to be returned. For further questions call (770) 670-2422.
- Quantity in commerce
- 101
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, IA, KY, LA, ME, MD, MA, MI, MO, MT, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TX, VT, VA and WA,. and the countries of Australia, Canada, New Zealand and United Kingdom.
- Product codes and lots
- Released builds 2.50.04D5 and higher
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1869-2014 | Class II | Terminated | Voluntary: Firm initiated |