510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1944-2014Class II

syngo Lab Data Manager, SMN 10800057, 10803189, and 10804573

Siemens Healthcare Diagnostics Inc. / initiated 2014-05-20 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of syngo Lab Data Manager, SMN 10800057, 10803189, and 10804573 on . FDA classified it as Class II and gives the reason as "Certain configuration parameters may cause the reference ranges to not hold results as expected and release them to the Laboratory Information System (LIS) that should have been held for manual review". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1944-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
syngo Lab Data Manager, SMN 10800057, 10803189, and 10804573.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Certain configuration parameters may cause the reference ranges to not hold results as expected and release them to the Laboratory Information System (LIS) that should have been held for manual review.
Root cause tag
design
FDA root cause
Device Design
Action
An urgent medical device correction notice, dated May 20, 2014, was sent to end users to provide instructions to avoid potential risks associated with the software defect, until changes can be made by a Siemens technical representative.
Quantity in commerce
336
Distribution
Worldwide Distribution. US Nationwide, Canada, Australia, Belgium, Czech Republic, Germany, Denmark, Finland, France, United Kingdom, Italy, Netherlands, Portugal, Slovakia, Singapore, Japan, and Spain.
Product codes and lots
syngo Lab Data Manager Application Software versions VA11B, VA12A, and VA12B

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1944-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1944-2014Class IITerminatedVoluntary: Firm initiated