Z-2133-2014Class III

Dimension Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids

Siemens Healthcare Diagnostics Inc. / initiated 2014-05-27 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids on . FDA classified it as Class III and gives the reason as "There is a misassembly issue with the photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2133-2014
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
There is a misassembly issue with the photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
An urgent field safety notice, dated May 2014, was sent to users to inform them of the issue and that a service representative would replace the misassembled photometer.
Quantity in commerce
15
Distribution
FL LA AR IA OH OH WI IL LA KS MT
Product codes and lots
DE271317, DE271320, DE271323, DE271324, DE271326, DE271327, DE271328, DE271332 DE271341, DE271343, DE271325, 12252589, 2004082898, 2004082899, 224719-AX

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2133-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2133-2014Class IIITerminatedVoluntary: Firm initiated