Dimension Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids on . FDA classified it as Class III and gives the reason as "There is a misassembly issue with the photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2133-2014
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- There is a misassembly issue with the photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- An urgent field safety notice, dated May 2014, was sent to users to inform them of the issue and that a service representative would replace the misassembled photometer.
- Quantity in commerce
- 15
- Distribution
- FL LA AR IA OH OH WI IL LA KS MT
- Product codes and lots
- DE271317, DE271320, DE271323, DE271324, DE271326, DE271327, DE271328, DE271332 DE271341, DE271343, DE271325, 12252589, 2004082898, 2004082899, 224719-AX
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2133-2014 | Class III | Terminated | Voluntary: Firm initiated |