Z-2002-2014Class II

3M Attest Auto-reader. Model numbers 390, 390G,and 490. These devices incubate the biological indicators...

3M Company / initiated 2014-06-09 / Terminated / root cause: labeling and UDI

3M Company - Health Care Business began a recall of 3M Attest Auto-reader. Model numbers 390, 390G,and 490. These devices incubate the biological indicators (BIs) to determine a positive or negative result for... on . FDA classified it as Class II and gives the reason as "Labeling on units shipped prior to May 16th, 2014 did not contain the statement "This product contains dry natural rubber" as required by the United States Food and Drug Administration (FDA) for medical devices. The...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2002-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
3M Company
Product
3M Attest Auto-reader. Model numbers 390, 390G,and 490. These devices incubate the biological indicators (BIs) to determine a positive or negative result for the sterilization cycle.
Product code
FRC Indicator, Biological Sterilization Process
Reason for recall
Labeling on units shipped prior to May 16th, 2014 did not contain the statement "This product contains dry natural rubber" as required by the United States Food and Drug Administration (FDA) for medical devices. The non-slip pads at the bottom of the unit contain dry natural rubber. One incident of an allergic reaction in a sensitized individual has been reported.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
Consignees were sent on 6/9/2014 a 3M "Urgent Medical Devices Correction" letter dated May 21, 2014. The letter described the problem and the product involved in the recall. The letter described what action is 3M taking and what action was required of the consignees. Consignees were sent an adhesive label and a response card. They requested consignees to adhere the label to a portion of the plastic covering each unit in a location where it is visible to the end-user. After that, they are requested to fill out and return the response card.
Quantity in commerce
5343 (2621 USA, 2722 OUS)
Distribution
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the countries of AUSTRALIA, CANADA, CHILE , COLOMBIA, COSTA RICA, SOUTH AFRICA, BRAZIL, GERMANY, ECUADOR, DUBAI, JAPAN, HONG KONG , CHINA, KOREA, MALAYSIA, PANAMA, SINGAPORE, VATICAN, and THAILAND.
Product codes and lots
All serial numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2002-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2002-2014Class IITerminatedVoluntary: Firm initiated