HydroFinity Hydrophilic guidewire;Nitinol Guidewire with Hydrophilic Coating, Sterile. 0.035 OD, Stiff Shaft...
Nitinol Devices and Components, Inc. began a recall of HydroFinity Hydrophilic guidewire;Nitinol Guidewire with Hydrophilic Coating, Sterile. 0.035 OD, Stiff Shaft, Angled Tip, 180 CM Length; Manufactured for... on . FDA classified it as Class I and gives the reason as "Reports of outer polymer jacket to the core wire being damaged during use on the HydroFinity Guidewire". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1819-2014
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Confluent Medical Technologies, Inc
- Product
- HydroFinity Hydrophilic guidewire;Nitinol Guidewire with Hydrophilic Coating, Sterile. 0.035 OD, Stiff Shaft, Angled Tip, 180 CM Length; Manufactured for Covidien, Plymouth, MN; Manufactured by: NDC Inc., Fremont, CA Manufactured in Costa Rica. The HydroFinity Hydrophilic guidewire facilitates the introduction and placement of catheters and interventional devices to the desired anatomical location during diagnostic or interventional procedures.
- Product code
- DQX Wire, Guide, Catheter
- Reason for recall
- Reports of outer polymer jacket to the core wire being damaged during use on the HydroFinity Guidewire.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Covidien sent an Urgent Product Recall letter dated May 30, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return any unused devices to Covidien, their account would be credited. Customers were also instructed to:: 1) Stop using product listed in this letter immediately 2) Segregate this product from other inventory 3) Fill out the reply (verification) form at the end of this letter If you do not have any product identified in this letter, please fax or email the completed form to Covidien at (877) 523-9109 or [email protected]. If you do have product, your sales representative will assist you in completing the verification form and arranging for return of the product. Please fax the completed form to Covidien (877) 523-9109 or email to [email protected]. 4) Your sales representative will be available to answer any questions regarding this recall and assist you in completing the verification form, returning product and addressing any account credits. This action is being conducted with the knowledge of the United States FDA and other regulatory authorities. Customers with questions were instructed to call (202) 310-5120. For questions regarding this recall call 202-310-5120. _
- Quantity in commerce
- 8485 in US, 9595 - ROW - total, all models
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Italy, France, Norway, Sweden, Germany, Belgium, United Kingdom and Canada.
- Product codes and lots
- Model HPSA35180- all lots.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1812-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1813-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1814-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1815-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1816-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1817-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1818-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1819-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1820-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1821-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1822-2014 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1823-2014 | Class I | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
