Z-1807-2014Class II
iGUIDE System
Elekta Limited / initiated 2014-05-29 / Terminated / root cause: software
Elekta, Inc. began a recall of iGUIDE System Product Usage: The intended use of the device is the control of accurate patient positioning with the assistance of a 30 Tracking System in a... on . FDA classified it as Class II and gives the reason as "A bug in the software prevents the iGUIDE software from logging off". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1807-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- iGUIDE System Product Usage: The intended use of the device is the control of accurate patient positioning with the assistance of a 30 Tracking System in a radiotherapy environment.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- A bug in the software prevents the iGUIDE software from logging off.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Elekta sent an Important Field Safety Notice 618 01 303 024 dated May 29, 2014 to affected customers. The letter identified the affected product, problem, potential clinical impact and actions to be taken. It also gives a recommendation how the risk can be mitigated - Users should only use the iGUIDE software when the key is in the on position and the Position Guard is active. Users are advised to restart the workstation if any error occurs to ensure the software is running properly. The notice also advises the customer that a patch will be released that corrects the behavior of the iGUIDE software. This solution will be implemented on all affected products when available. For questions contact your local Elekta office.
- Quantity in commerce
- 224
- Distribution
- Worldwide Distribution - USA (nationwide) in the states of AZ, CA, CT, HI, IL, IA, KY, MD, MA, MI, MO, NH, NJ, NC, ND, OH, OK, OR, PA, TX, VA, WA, WI, and the countries of Algeria, Australia, Austria, Brazil, Canada, China, Colombia, Denmark, Finland, France, Germany, Hong Kong, India, Italy, Japan, Kazakhstan, Mexico, Netherlands, New Zealand, Peru, Poland, Russia, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom and Vietnam
- Product codes and lots
- Software Versions 2.0.0, 2.0.1 and 2.0.2