Z-1809-2014Class II

Model Number(s): PB840 Ventilator (4-840120DIUU-XX). The product is a component of the Puritan Bennett 840...

Medtronic, Inc. / initiated 2014-06-04 / Terminated / root cause: design

Nellcor Puritan Bennett Inc. (dba Covidien LP) began a recall of Model Number(s): PB840 Ventilator (4-840120DIUU-XX). The product is a component of the Puritan Bennett 840 Ventilator: The Backlight Inverter Printed Circuit... on . FDA classified it as Class II and gives the reason as "In the case of a loss of GUI display due to a Backlight Inverter PCBA failure, the ventilator continues to provide uninterrupted ventilatory support at the programmed settings for the patient. However, there is a loss...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1809-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Model Number(s): PB840 Ventilator (4-840120DIUU-XX). The product is a component of the Puritan Bennett 840 Ventilator: The Backlight Inverter Printed Circuit Board (BLI PCBA) for the 9.4 Display part number 4-079056-00. The PB840 Ventilator System is intended to provide invasive or non-invasive ventilator support and monitoring for infant, pediatric, and adult patients with respiratory failure or respiratory insufficiency. It is used in hospitals and healthcare facilities
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
In the case of a loss of GUI display due to a Backlight Inverter PCBA failure, the ventilator continues to provide uninterrupted ventilatory support at the programmed settings for the patient. However, there is a loss of display and thus there is a necessity to move the patient to another ventilator.
Root cause tag
design
FDA root cause
Component design/selection
Action
Covidien will be issuing the Product Notification letter in 2014. Customers are requested to forward the letter to anyone they may have transferred or assigned for use one of the 9.4display configuration ventilators. Service engineers will be replacing the BLI PCBA with a new design containing the 1.0 ampere fuse and then destroying the old PCBA. Ventilators replaced with newer models under the commercial program will be returned to Covidien for disposal. Should you have any questions regarding this letter or to report any issues with the Puritan Bennett" 840 ventilator contact our Technical Support Department at 1-800-255-6774, option 4, then option 1, to ensure proper device reporting procedures are followed. Thank you for your prompt attention to this matter.
Quantity in commerce
7371
Distribution
Worldwide Distribution - USA (nationwide) including the states: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and District of Columbia., and the countries of Canada and Japan.
Product codes and lots
Display part number 4-079056-00; serial number 4200000001 to 3510012377 that have a 9.4 display. Lots 1032H,1041H,1042H,1043H,1008J,1009J,1010J,1018K,1019K,1020K,1021K,1022K,1023K,1030L,1031L,1032L,1033L,1034L,1035L,

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1809-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1809-2014Class IITerminatedVoluntary: Firm initiated