Z-1979-2014Class II

CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs...

Nihon Kohden America Inc / initiated 2014-06-03 / Terminated / root cause: software

Nihon Kohden America Inc began a recall of CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients on . FDA classified it as Class II and gives the reason as "Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1979-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Nihon Kohden America Inc
Product
CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.
Root cause tag
software
FDA root cause
Software design
Action
Nihon Kohden sent an Urgent Medical Device Safety Notification and Field Correction letter dated June 3, 2014 to affected customers. Recall notification letter, dated June 3, 2014 was hand-delivered to site by Nihon personnel. The letter provided the following information: product description with codes and software version, reason for recall, field correction information and actions to be taken, MedWatch reporting information, and contact information for recall.
Quantity in commerce
430
Distribution
Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.
Product codes and lots
Model Number: CNS-6201  Software Version 02-26 only. Serial Numbers: 00485,  00488,  00489,  00490,  00491,  00361,  00429,  00430, 00432,  00436,  00443,  00447,  00478,  00486,  00546,  00564,  00573,  00574,  00575,  00576,  00592,  00626, 00638, 00643,  00651,  00674,  00675,  00677,  00678, 00679, 00687.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1979-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1979-2014Class IITerminatedVoluntary: Firm initiated