510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1973-2014Class II

Cautery QRG for the da Vinci Standard Surgical System; da Vinci S System and da Vinci Si System . Intuitive...

Intuitive Surgical, Inc. / initiated 2014-05-21 / Terminated / root cause: design

Intuitive Surgical, Inc. began a recall of Cautery QRG for the da Vinci Standard Surgical System; da Vinci S System and da Vinci Si System . Intuitive Surgical Endoscopic Instrument Control System on . FDA classified it as Class II and gives the reason as "Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1973-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intuitive Surgical, Inc.
Product
Cautery QRG for the da Vinci Standard Surgical System; da Vinci S System and da Vinci Si System . Intuitive Surgical Endoscopic Instrument Control System.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
Multiple updates to User Manuals, Instructions for Use (IFU), Quick Reference Guides (QRG), and Instrument Release Kits (IRK) for use with the da Vinci Surgical System, instruments and accessories. Removal of the Emergency Grip Release Wrench, part number 710142.
Root cause tag
design
FDA root cause
Device Design
Action
Intuitive Surgical sent a Field Safety Notice dated June 16, 2014, to all consignees affected by changes in labeling, IFU User manuals and QRG's. Consignees were instructed to take the following actions: Ensure all appropriate personnel are fully informed of this Notification. Forward this letter to your Risk Manager, OR Director, Purchasing Manager, Biomedical Engineering staff and members of your medical staff who perform da Vinci surgery procedures. 2. Use Attachment C to identify the labeling that has been updated. 3. No manuals or products should be returned. Discard old versions of the User Manuals, IFUs, and QRGs for which you have received updated versions. 4. Discard all versions of the Emergency Grip Release Wrench (Part Number 710142). Order additional IRKs from Intuitive Surgical as needed. 5. Complete and return the attached Acknowledgement Form to acknowledge that you have received the updated Labeling, IRKs, in-service and have discarded all previous versions of Labeling and the emergency grip release wrench. Please return the completed Acknowledgment Form within two days of the in-service. 6. Retain a copy of this Notice and the Acknowledgement Form for your records. If you need further information or support concerning this recall, please contact your Intuitive Surgical Representative or Intuitive Surgical Customer Service at the numbers listed below: North America: 800-876-1310 Option 3 (6 am to 5pm PST) For questions regarding this recall call 408-523-2602.
Quantity in commerce
All distributed prior to June 2014
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Great Britain, Greece, Guadeloupe, India, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Norway, Pakistan, Philippines, Indonesia, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovak Republic, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uruguay, Venezuela, and Vietnam.
Product codes and lots
Part number 551992.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1973-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1964-2014Class IITerminatedVoluntary: Firm initiated
Z-1965-2014Class IITerminatedVoluntary: Firm initiated
Z-1966-2014Class IITerminatedVoluntary: Firm initiated
Z-1967-2014Class IITerminatedVoluntary: Firm initiated
Z-1968-2014Class IITerminatedVoluntary: Firm initiated
Z-1969-2014Class IITerminatedVoluntary: Firm initiated
Z-1970-2014Class IITerminatedVoluntary: Firm initiated
Z-1971-2014Class IITerminatedVoluntary: Firm initiated
Z-1972-2014Class IITerminatedVoluntary: Firm initiated
Z-1973-2014Class IITerminatedVoluntary: Firm initiated
Z-1974-2014Class IITerminatedVoluntary: Firm initiated
Z-1975-2014Class IITerminatedVoluntary: Firm initiated