Z-1853-2014Class II
RT-5100 Refractor and RT-3100 Refractor; Distributed by MARCO Opthalmic: Manufactured by NIDEK CO., Ltd...
Nidek Incorporated / initiated 2014-06-09 / Terminated
Nidek Inc began a recall of RT-5100 Refractor and RT-3100 Refractor; Distributed by MARCO Opthalmic: Manufactured by NIDEK CO., Ltd., Aichi, Japan. Instrument provides means of... on . FDA classified it as Class II and gives the reason as "Near Point Chart Arm of RT 5100 and RT-3100 Refractors may lower spontaneously with potential for patient injury". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1853-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nidek Incorporated
- Product
- RT-5100 Refractor and RT-3100 Refractor; Distributed by MARCO Opthalmic: Manufactured by NIDEK CO., Ltd., Aichi, Japan. Instrument provides means of positioning spherical and cylindrical lenses, prisms and other optical devices in front of a subject's eyes for purpose of determining refractive error and binocular functions.
- Product code
- HKN Refractor, Manual, Non-Powered, Including Phoropter
- Reason for recall
- Near Point Chart Arm of RT 5100 and RT-3100 Refractors may lower spontaneously with potential for patient injury.
- FDA root cause
- Use error
- Action
- Recall/Field Correction letters were sent to doctors on June 9, 2014 with a preventive maintenance instruction to be performed on the device by user. Also included are the necessary tools to perform actions. Customers are to complete and return the Recall Response Card. Questions are to be directed to Customer Service. Contact Information: Customer Service: 1-800-891-1772 E-mail: RT-5100 [email protected]
- Quantity in commerce
- ~4,285 units RT-5100; 167 units of RT-3100
- Distribution
- US distribution only.
- Product codes and lots
- RT-5100 and RT-3100, various serial numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1853-2014 | Class II | Terminated | Voluntary: Firm initiated |