Z-1977-2014Class II
BD FACS Canto and associated Fluidics Cart
Becton, Dickinson and Company, BD Biosciences / initiated 2014-06-06 / Terminated / root cause: manufacturing process
BD Biosciences, Systems & Reagents began a recall of BD FACS Canto and associated Fluidics Cart Product Usage: Flow cytometer for use in In vitro Diagnostics on . FDA classified it as Class II and gives the reason as "Laser cooling fans and internal fan of the associated fluidics cart are not operational". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1977-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Product
- BD FACS Canto and associated Fluidics Cart Product Usage: Flow cytometer for use in In vitro Diagnostics.
- Product code
- OYE Flow Cytometric Reagents And Accessories.
- Reason for recall
- Laser cooling fans and internal fan of the associated fluidics cart are not operational.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- BD sent an Important Product Recall notification letters dated June 2014 to affected US customers by trackable mail. International affiliates were notified on June 6, 2014. The letter identified the affected product, problem and the actions to be taken. A Field Change notification has been established and a BD field Service Engineer will coordinate with each account to perform the necessary corrections. Questions or concerns may be directed to BD Customer Support at BD Biosciences, prompt 2. Out of US customers are to contact local BD rep with questions.
- Quantity in commerce
- 35 installed devices
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, OH, RI, GA, FL, MD, NY, WV, NJ, Ma, CT, IA, NM, MN and the countries of China and Belgium.
- Product codes and lots
- FACS Canto: model number 657338; Associates Fluidic Cart (catalog number 6657660)- IVD; Serial numbers: V65733800001, V65733800002, V65733800003, V65733800004 V65733800005, V65733800006, V65733800007, V65733800008, V65733800009, V657338000010, V657338000011, V657338000012, V657338000013, V657338000014, V657338000017, V657338000018, V657338000019, V657338000020, V657338000021, V657338000022, V657338000023, V657338000024, V657338000025, V657338000026, V657338000028, V657338000029, V657338000031, V657338000032, V657338000033, V657338000034, V65733800035, V65733800036, V65733800037, V65733800038, V65733800039, V657338000040 V657338000043, V657338000044, V657338000045, V657338000046, V657338000047, V657338000048, V657338000049, V657338000050, V657338000051, V657338000052, V657338000053, V657338000054, V657338000055, V657338000056, V657338000057, V657338000058, V657338000060.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1977-2014 | Class II | Terminated | Voluntary: Firm initiated |