Z-1983-2014Class II

Anspach Pneumatic and Electric Motor Systems. Intended for cutting and shaping bone, including the spine and...

The Anspach Effort, Inc. / initiated 2014-05-08 / Terminated / root cause: manufacturing process

The Anspach Effort, Inc. began a recall of Anspach Pneumatic and Electric Motor Systems. Intended for cutting and shaping bone, including the spine and cranium on . FDA classified it as Class II and gives the reason as "Several attachments and handpieces failed to meet the expected sterility assurance level (SAL) when sterilized by the method that is specified in the current Directions for Use (DFU) provided with the device". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1983-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
The Anspach Effort, Inc.
Product
Anspach Pneumatic and Electric Motor Systems. Intended for cutting and shaping bone, including the spine and cranium.
Product code
HSZ Instrument, Surgical, Orthopedic, Pneumatic Powered & Accessory/Attachment
Reason for recall
Several attachments and handpieces failed to meet the expected sterility assurance level (SAL) when sterilized by the method that is specified in the current Directions for Use (DFU) provided with the device.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Customers were notified 05/08/2014 by a Return Receipt Requested letter from Anspach with an URGENT: Medical Device Labeling Correction announcement, REGARDING: STERRAD 100S Hydrogen Peroxide Gas Plasma Technology no longer recommended for sterilization of Anspach products. The letter identified the problem and actions to be taken by the customer. Customers were to complete and return the attached reply form. If the products were distributed further, then those customers should have been notified as well. Customers were to contact Anspach Customer Support at 1-800-327-6887 if they want copies of the revised labeling.
Quantity in commerce
115,988 units
Distribution
Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX,UT, VA, VT, WA, WI, WV, WY, and Hawaii, and the territories of Puerto Rico, US Virgin Island and Guam; and, the countries of Argentina, Australia, Austria, Belgium, Brazil, Cameroon, Canada, Chile, Costa Rica, Czech Republic, Dominican Republic, Egypt, Finland, France, Germany, Grand Cayman, Hong Kong, India, Israel, Jamaica, Japan, Korea, Lithuania, Malaysia, Mexico, Nassau, Netherlands, Nigeria, Palestine, Panama, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Product codes and lots
DFU:  18-0047: Revisions A & B  18-0074: Revision A; and  18-0105: Revisions A.  Operating Manuals:  19-0010: Revisions A, and  19-0013: Revisions A, B, & C.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1983-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1983-2014Class IITerminatedVoluntary: Firm initiated