Z-1827-2014Class II

Philips MR systems: Trolley Variable Height IRF (for Ingenia 1.5T & 3.0T products)

Philips North America LLC / initiated 2014-05-15 / Terminated / root cause: design

Philips Medical Systems, Inc. began a recall of Philips MR systems: Trolley Variable Height IRF (for Ingenia 1.5T & 3.0T products) Product Usage: Diagnostic MR System. The Height Adjustable FlexTrak trolley... on . FDA classified it as Class II and gives the reason as "Ingenia customers have experienced clamping of the foot under the central column of the Height Adjustable Flex Trak trolley and the Trolley Variable Height IRF". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1827-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips MR systems: Trolley Variable Height IRF (for Ingenia 1.5T & 3.0T products) Product Usage: Diagnostic MR System. The Height Adjustable FlexTrak trolley and Trolley Variable Height IRF is an optional device designed to prepare the patient outside the examination room to enhance workflow and make additional non-magnetic hospital beds unnecessary.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Ingenia customers have experienced clamping of the foot under the central column of the Height Adjustable Flex Trak trolley and the Trolley Variable Height IRF.
Root cause tag
design
FDA root cause
Device Design
Action
Philips sent an Urgent Field Safety Notice (FSN78100414) dated May 2014 via certified mail to US and Canadian consignees. Philips Key Market representatives will distribute the letter internationally. The letter identified the affected product, explains the problem, actions that should be taken in order to prevent risks for users, and the actions planned by Philips to correct the problem. On May 31, 2014, Philips released a Field Change Order that requires service engineers to replace the front and rear covers of the Ingenia MR FlexTrak Trolley with new covers with protection bumper that prevents foot clamping. For further information and support concerning this issue, consignees should contact their local Philips representative at 1-800-722-9377.
Quantity in commerce
5
Distribution
Worldwide Distribution - US (nationwide), Canada, Mexico Australia, Austria, Germany, Belarus, Bulgaria, Brazil, China, Denmark, Estonia, Finland, France, Germany, Greece, Italy, Jordan, Korea, Republic of Korea, Netherlands, New Zealand, Norway, Oman, Pakistan, Russian Federation, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Product codes and lots
Model Number: 9896 030 1929.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1827-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1826-2014Class IITerminatedVoluntary: Firm initiated
Z-1827-2014Class IITerminatedVoluntary: Firm initiated