Z-1978-2014Class II
Stryker Orthopaedics Triathlon Tritanium Patella Inserter, . NON-STERILE
Howmedica Osteonics Corp / initiated 2014-05-01 / Terminated / root cause: manufacturing process
Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Triathlon Tritanium Patella Inserter, . NON-STERILE Product Usage: The Triathlon Tritanium Patella Inserter is part of the Tritanium Knee... on . FDA classified it as Class II and gives the reason as "Tritanium Patella Inserter Instrument fracture during implantation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1978-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Orthopaedics Triathlon Tritanium Patella Inserter, . NON-STERILE Product Usage: The Triathlon Tritanium Patella Inserter is part of the Tritanium Knee System Instrument Set. It is used to implant the Tritanium Patella after the patellar bone has been prepared using the existing Triathlon Patella Resection Guide, Metal Backed Patella Drill and Patella Peg Drill Templates as per the Triathlon Tritanium Surgical Protocol.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Tritanium Patella Inserter Instrument fracture during implantation. .
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Stryker Orthopaedics issued an "Urgent Medical Device Recall Notification" letters/Notification Acknowledgement forms dated May 7, 2014 to their customers via Fed-Ex on May 7, 2014. The notification informs the customers of the issue(s) with the affected product; potential hazards; and how to identify and handle any affected product. Customers were instructed to return any affected product to Stryker Orthopaedics and fill out and return the enclosed Notification Acknowledgement form. For questions call (1-201-972-2100).
- Quantity in commerce
- 22 Units
- Distribution
- US Nationwide Distribution - in the states of Arizona, Florida, Kentucky, Louisiana and New Jersey.
- Product codes and lots
- Catalog No. 6541-3-530. Lot Codes: CTD2778 and CTD2779.