Z-1978-2014Class II

Stryker Orthopaedics Triathlon Tritanium Patella Inserter, . NON-STERILE

Howmedica Osteonics Corp / initiated 2014-05-01 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Triathlon Tritanium Patella Inserter, . NON-STERILE Product Usage: The Triathlon Tritanium Patella Inserter is part of the Tritanium Knee... on . FDA classified it as Class II and gives the reason as "Tritanium Patella Inserter Instrument fracture during implantation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1978-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Orthopaedics Triathlon Tritanium Patella Inserter, . NON-STERILE Product Usage: The Triathlon Tritanium Patella Inserter is part of the Tritanium Knee System Instrument Set. It is used to implant the Tritanium Patella after the patellar bone has been prepared using the existing Triathlon Patella Resection Guide, Metal Backed Patella Drill and Patella Peg Drill Templates as per the Triathlon Tritanium Surgical Protocol.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Tritanium Patella Inserter Instrument fracture during implantation. .
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Stryker Orthopaedics issued an "Urgent Medical Device Recall Notification" letters/Notification Acknowledgement forms dated May 7, 2014 to their customers via Fed-Ex on May 7, 2014. The notification informs the customers of the issue(s) with the affected product; potential hazards; and how to identify and handle any affected product. Customers were instructed to return any affected product to Stryker Orthopaedics and fill out and return the enclosed Notification Acknowledgement form. For questions call (1-201-972-2100).
Quantity in commerce
22 Units
Distribution
US Nationwide Distribution - in the states of Arizona, Florida, Kentucky, Louisiana and New Jersey.
Product codes and lots
Catalog No. 6541-3-530. Lot Codes: CTD2778 and CTD2779.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1978-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1978-2014Class IITerminatedVoluntary: Firm initiated