Z-2174-2014Class II

Siemens Linear Accelerators of type ARTISTE, ONCOR, and PRIMUS with Automatic Sequenced Cancer Treatment...

Siemens Medical Solutions USA, Inc. / initiated 2014-05-28 / Terminated

Siemens Medical Solutions USA, Inc began a recall of Siemens Linear Accelerators of type ARTISTE, ONCOR, and PRIMUS with Automatic Sequenced Cancer Treatment Delivery Option on . FDA classified it as Class II and gives the reason as "A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2174-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Linear Accelerators of type ARTISTE, ONCOR, and PRIMUS with Automatic Sequenced Cancer Treatment Delivery Option.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to lead to a collision with the patient.
FDA root cause
Use error
Action
An Urgent Medical Device Correction Field Safety Notice Letter was sent to end users on/about June 2014 which described the product, problem, and actions to be taken to avoid encountering the safety risk to patients and users.
Quantity in commerce
160
Distribution
Nationwide and ROW, including Canada and Mexico.
Product codes and lots
material numbers: 01940035, 04504200, 05672977, 05693908, 05857912, 05857920, 08139789, 08319758, 08490005, 08515520, 09401654, 09401746, 09822693, with serial numbers below 3094.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2174-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2174-2014Class IITerminatedVoluntary: Firm initiated