Z-1990-2014Class II

MicroScan WalkAway-40 plus Instrument and MicroScan Walkaway-96 plus instrument access door hinge. Siemens...

Siemens Healthcare Diagnostics Inc. / initiated 2014-06-02 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of MicroScan WalkAway-40 plus Instrument and MicroScan Walkaway-96 plus instrument access door hinge. Siemens Healthcare Diagnostics, Inc. in vitro diagnostic on . FDA classified it as Class II and gives the reason as "Springs contained in the access door hinge assembly on the Walk Away Plus instruments are becoming detached causing premature failure of the hinge assembly and a potential safety issue for the customers". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1990-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
MicroScan WalkAway-40 plus Instrument and MicroScan Walkaway-96 plus instrument access door hinge. Siemens Healthcare Diagnostics, Inc. in vitro diagnostic
Product code
LRG Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems
Reason for recall
Springs contained in the access door hinge assembly on the Walk Away Plus instruments are becoming detached causing premature failure of the hinge assembly and a potential safety issue for the customers.
Root cause tag
design
FDA root cause
Component design/selection
Action
Urgent Medical Device Recall notification letter was sent to all affected US customers on June 2nd by Federal Express. OUS notifications were also sent to regional Siemens organizations who distributed in-country notifications and relevant notifications.
Quantity in commerce
WA40: 560 US, 131 OUS, 9 already repaired; WA96: 81 US, 189 OUS, 57 repaired.
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Algeria Argentina Brazil Canada China Chile Columbia France Indonesia Libya Georgia Germany India Italy Japan Kazakhstan Lithuania Oman Pakistan Peru Poland Portugal Republic of Korea Romania Russia Saudi Arabia Spain Turkey.
Product codes and lots
WalkAway 40 - Material number: 10444853; catalog number: B1018-283; WalkAway 96 - Material number: 1044854, Catalog number B1018-284.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1990-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1990-2014Class IITerminatedVoluntary: Firm initiated