Z-1927-2014Class II

Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot...

Siemens Healthcare Diagnostics Inc. / initiated 2014-06-02 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc began a recall of Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot number: 454778 Kit Catalog Number: 112219 SMN... on . FDA classified it as Class II and gives the reason as "Swollen acid bottles due to a trace metal contamination in the Acid Reagent bottles". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1927-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot number: 454778 Kit Catalog Number: 112219 SMN: 10310026 Intended Use: In vitro diagnostic use. Acid Regent is a component of the ADVIA Centaur and ADVIA Centaur XP instruments
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Swollen acid bottles due to a trace metal contamination in the Acid Reagent bottles
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Siemens Healthcare Diagnostics issued Urgent Medical Device Recall (UMDR) to customers in the United States on June 02, 2014 and a Urgent Field Safety Notice (UFSN) issued outside the United States on June 02, 2014. These notices inform the customer of the potential for swollen acid bottles due to confirmed metal contamination. Customers with alternative lots in inventory were instructed to transition to the alternative lot upon receipt of the communication. Customers without an alternative lot of acid in inventory were informed that Siemens proactively shipped an alternative lot to their laboratory and they should transition to that lot upon its receipt. Siemens also informed customers without an alternative lot that they may continue to run and report patient results using the affected lot of acid. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
Quantity in commerce
3,839 kits
Distribution
Worldwide Distribution - US Nationwide and the countries of Canada, Argentina, China, Colombia, Korea, Malyasia, New Zealand, Mexico, Guatemala, and Singapore.
Product codes and lots
Kit Lot Number: 454779 Exp. Date: 28 February, 2015

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1927-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1927-2014Class IITerminatedVoluntary: Firm initiated