Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot...
Siemens Healthcare Diagnostics, Inc began a recall of Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot number: 454778 Kit Catalog Number: 112219 SMN... on . FDA classified it as Class II and gives the reason as "Swollen acid bottles due to a trace metal contamination in the Acid Reagent bottles". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1927-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot number: 454778 Kit Catalog Number: 112219 SMN: 10310026 Intended Use: In vitro diagnostic use. Acid Regent is a component of the ADVIA Centaur and ADVIA Centaur XP instruments
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Swollen acid bottles due to a trace metal contamination in the Acid Reagent bottles
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- Siemens Healthcare Diagnostics issued Urgent Medical Device Recall (UMDR) to customers in the United States on June 02, 2014 and a Urgent Field Safety Notice (UFSN) issued outside the United States on June 02, 2014. These notices inform the customer of the potential for swollen acid bottles due to confirmed metal contamination. Customers with alternative lots in inventory were instructed to transition to the alternative lot upon receipt of the communication. Customers without an alternative lot of acid in inventory were informed that Siemens proactively shipped an alternative lot to their laboratory and they should transition to that lot upon its receipt. Siemens also informed customers without an alternative lot that they may continue to run and report patient results using the affected lot of acid. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
- Quantity in commerce
- 3,839 kits
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Canada, Argentina, China, Colombia, Korea, Malyasia, New Zealand, Mexico, Guatemala, and Singapore.
- Product codes and lots
- Kit Lot Number: 454779 Exp. Date: 28 February, 2015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1927-2014 | Class II | Terminated | Voluntary: Firm initiated |