510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2032-2014Class II

Zimmer EDI PERIARTICULAR PLATING SYSTEM CANCELLOUS BONE SCREW, FULLY THREADED, HEX HEAD 4MM DIAMETER, 14 MM...

Zimmer Manufacturing B.V. / initiated 2014-06-12 / Terminated / root cause: manufacturing process

Zimmer Manufacturing B.V. began a recall of Zimmer EDI PERIARTICULAR PLATING SYSTEM CANCELLOUS BONE SCREW, FULLY THREADED, HEX HEAD 4MM DIAMETER, 14 MM LENGTH, and EDI 4.0 MM CANCELLOUS SCREW, FULLY... on . FDA classified it as Class II and gives the reason as "Two lots of screws were commingled. Screws with etching and machining for lot number 62629432 were packaged in bags labeled for lot number 62628759; and screws with etching and machining for lot number 62628759 were...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2032-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer Manufacturing B.V.
Product
Zimmer EDI PERIARTICULAR PLATING SYSTEM CANCELLOUS BONE SCREW, FULLY THREADED, HEX HEAD 4MM DIAMETER, 14 MM LENGTH, and EDI 4.0 MM CANCELLOUS SCREW, FULLY THREADED, 20 MM LENGTH. Temporary internal fixation devices, designed to stabilize fractures during the normal healing process.
Product code
HWC Screw, Fixation, Bone
Reason for recall
Two lots of screws were commingled. Screws with etching and machining for lot number 62629432 were packaged in bags labeled for lot number 62628759; and screws with etching and machining for lot number 62628759 were packaged in bags labeled for lot number 62629432. The affected product was distributed between 3/6/14 and 4/21/14.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Zimmer sent an Urgent Medical Device Recall Lot specific letter dated June 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed: 1. Please locate, quarantine, and return the affected product (part number 00-4840-020-00 lot number 62629432 and part number 00-2347-022-14 lot number 62628759) from all inventories. 2. Please note that any hospitals that received direct shipments of this product from Zimmer will be notified via Fed Ex. 3. Please review the list of hospital and surgeons to be provided to your territory via email. 4. Provide the name, address, and phone number of any additional hospitals or surgeons that may have received or used the affected product, or confirm that the list provided is complete. 5. Complete and return Attachment 1. 6. Ensure that a copy of the Risk Managers and Surgeons letter is left with the facilities from which product is removed by providing them with a copy of the letter, if they are not identified as a facility receiving a letter directly from Zimmer. 7. If after reviewing this notification you have questions or concerns, please contact Zimmer at 1-877-946-2761.
Quantity in commerce
496
Distribution
Worldwide Distribution - USA (nationwide) and the countries of India, Japan, China, Australia and Canada.
Product codes and lots
Lot No. 62629432 Lot No. 62628759

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2032-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2032-2014Class IITerminatedVoluntary: Firm initiated