Z-2059-2014Class II

V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human...

Datascope Corp. / initiated 2014-05-19 / Terminated / root cause: software

Mindray DS USA, Inc. dba Mindray North America began a recall of V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human physiological parameters. The V-Series Monitor also has... on . FDA classified it as Class II and gives the reason as "ECG interpretation feature issues. False measurements reported when the algorithm cannot determine P wave or QRS axis. The Median Complex Report displays the representative beat 25% larger than actual signal input. When...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2059-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human physiological parameters. The V-Series Monitor also has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
ECG interpretation feature issues. False measurements reported when the algorithm cannot determine P wave or QRS axis. The Median Complex Report displays the representative beat 25% larger than actual signal input. When a patient date of birth is imported from an external source, the ECG interpretation algorithm uses the default age of 50, and interpretive statements may be incorrect.
Root cause tag
software
FDA root cause
Software design
Action
Mindray issued "Urgent: V-Series Monitoring System Corrective Action" notices dated May 19, 2014, to customers that received the affected product via certified mail with return receipt verification. The notification informs the customer of the issues with the affected product and how to identify the affected product. The notification recommends that customers discontinue use of the 12-lead ECG interpretation feature until a software correction is available (anticipate August 2014). A point of contact is provided to customers in case they have questions. For further questions please call 1-800-288-2121.
Quantity in commerce
113 units (57 United States; 56 International)
Distribution
Worldwide Distribution - US including the states of Louisiana, New Jersey, Montana, South Carolina and Wyoming.
Product codes and lots
Part Number (P/N): 0998-00-1800-101 and 0998-00-1800-201 with software version 2.4.1.18 in use with 12-lead ECG Module, P/N:0998-00-1804-01 and 0998-00-1804-02.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2059-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2059-2014Class IITerminatedVoluntary: Firm initiated