Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The...
Remel Inc began a recall of Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The firm name on the label is Remel Inc., Lenexa, KS on . FDA classified it as Class II and gives the reason as "Bottles did not provide positive signals within the time frame of the quality control specification for one of two strains of Streptococcus pneumoniae". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2057-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Remel Inc
- Product
- Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The firm name on the label is Remel Inc., Lenexa, KS.
- Product code
- MDB System, Blood Culturing
- Reason for recall
- Bottles did not provide positive signals within the time frame of the quality control specification for one of two strains of Streptococcus pneumoniae.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The recalling firm made phone calls to their customers beginning 6/11/14 with follow-up letters dated 6/12/14 being issued via regular mail requesting the affected product be destroyed. Additional customers were later identified and additional phone calls were made on 6/16/14 and recall letters dated 6/16/14 were issued via regular mail.
- Quantity in commerce
- 76/50-bottle boxes
- Distribution
- Distribution was made to CO, IL, IN, NE, NJ, PA, SD, and a government facility located in VA.
- Product codes and lots
- Segments 1, 5, and 6 only of Lot 492571, Exp. 2015-10
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2057-2014 | Class II | Terminated | Voluntary: Firm initiated |