Z-2057-2014Class II

Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The...

Remel Inc / initiated 2014-06-11 / Terminated / root cause: manufacturing process

Remel Inc began a recall of Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The firm name on the label is Remel Inc., Lenexa, KS on . FDA classified it as Class II and gives the reason as "Bottles did not provide positive signals within the time frame of the quality control specification for one of two strains of Streptococcus pneumoniae". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2057-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Remel Inc
Product
Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The firm name on the label is Remel Inc., Lenexa, KS.
Product code
MDB System, Blood Culturing
Reason for recall
Bottles did not provide positive signals within the time frame of the quality control specification for one of two strains of Streptococcus pneumoniae.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The recalling firm made phone calls to their customers beginning 6/11/14 with follow-up letters dated 6/12/14 being issued via regular mail requesting the affected product be destroyed. Additional customers were later identified and additional phone calls were made on 6/16/14 and recall letters dated 6/16/14 were issued via regular mail.
Quantity in commerce
76/50-bottle boxes
Distribution
Distribution was made to CO, IL, IN, NE, NJ, PA, SD, and a government facility located in VA.
Product codes and lots
Segments 1, 5, and 6 only of Lot 492571, Exp. 2015-10

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2057-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2057-2014Class IITerminatedVoluntary: Firm initiated