Z-1946-2014Class II
Name: Pasteur Pipets, used by the embryologist in the IVF Laboratory. Model numbers: PP-5.75-90 PP-9-90...
CooperSurgical, Inc. / initiated 2014-06-20 / Terminated / root cause: packaging
ORIGIO, Inc began a recall of Name: Pasteur Pipets, used by the embryologist in the IVF Laboratory. Model numbers: PP-5.75-90 PP-9-90 PP-5.75-90PL PP-9-90PL on . FDA classified it as Class II and gives the reason as "Identification of packaging breach of the nylon pouch sterile barrier for the 3-pack configuration". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1946-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CooperSurgical, Inc.
- Product
- Name: Pasteur Pipets, used by the embryologist in the IVF Laboratory. Model numbers: PP-5.75-90 PP-9-90 PP-5.75-90PL PP-9-90PL
- Product code
- MQK Labware, Assisted Reproduction
- Reason for recall
- Identification of packaging breach of the nylon pouch sterile barrier for the 3-pack configuration
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Field Safety Corrective Action / Urgent Product Recall letter dated June, 2014.
- Quantity in commerce
- Total number of Boxes Distributed 32,760. US 4,093 / Canada 1,156 / EU & Other 27,511
- Distribution
- Worldwide Distribution: US Nationwide, Canada, Australia, Barbados, Chile, China, Colombia, Denmark, Ecuador, France, Germany, Great Britain, Greece, India, Italy, Japan, Jordan, Malaysia, Netherlands, Peru, Russia, Singapore, South Africa, Spain, Taiwan, Turkey, and United Arab Emirates.
- Product codes and lots
- CODE: Model #s: PP-5.75-90, PP-9-90, PP-5.75-90PL, PP-9-90PL. Lots with labeled expiry date of May 2, 2015, or later.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1946-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY