Enhancement Medical expression. model # EXP-150. Injectable Gel 1.5cc Syringe with hyaluronic acid...
Enhancement Medical, LLC began a recall of Enhancement Medical expression. model # EXP-150. Injectable Gel 1.5cc Syringe with hyaluronic acid concentration of 20 mg/ml. Product Usage: Expression is... on . FDA classified it as Class I and gives the reason as "Manufacturer made a change in the production process that resulted in a change in final gel weight. RECALL EXPANDED 7/8/2014 Firm expanded their recall to include all lots of product". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2135-2014
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Enhancement Medical, LLC
- Product
- Enhancement Medical expression. model # EXP-150. Injectable Gel 1.5cc Syringe with hyaluronic acid concentration of 20 mg/ml. Product Usage: Expression is labelled as an Intranasal Splint, intended to minimize bleeding and edema and to prevent adhesions between the septum and the nasal cavity. However, Expression is currently being used off-label as a subcutaneously injected dermal filler.
- Product code
- LMH Implant, Dermal, For Aesthetic Use
- Reason for recall
- Manufacturer made a change in the production process that resulted in a change in final gel weight. RECALL EXPANDED 7/8/2014 Firm expanded their recall to include all lots of product.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Enhancement Medical, LLC called all Independent Sales Representatives and then provided each representative a list of consignees with impacted product and instructed the Representatives to disclose low gel weight issue and potential for common treatment responses. Representatives were instructed to exchange the affected product. EXPANDED Enhancement Medical sent an Urgent Medical Device Recall letter beginning July 8, 2014 to all consignees whom received product. The letter identified all lots as being recalled, the reason for the recall, recommended that Expression not be injected subcutaneously, and asked for product to no longer be used and be returned to the recalling firm. An acknowledgement Form was asked to be mailed, e-mailed or faxed back to the firm. Customers with questions were instructed to call 414-918-4280.
- Quantity in commerce
- 17,875
- Distribution
- Worldwide Distribution - USA (nationwide including Puerto Rico) and Internationally to Canada.
- Product codes and lots
- Lot numbers to follow: Z36-074, Q36-136, Q36-137, Q36-138, Q36-139, Q36-140, R36-168, R36-169, R36-170, R36-171, R36-172, S36-189, S36-190, S36-191, S36-192, S36-194, S36-202, S36-203. RECALL EXPANDED 7-8-2014 to include ALL LOTS.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2135-2014 | Class I | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.