SmartSite Low Sorbing Infusion Set, Model Codes: 10014855A, 10015861A, 10061661A 14035457 14035485 14046894...
CareFusion 303, Inc. began a recall of SmartSite Low Sorbing Infusion Set, Model Codes: 10014855A, 10015861A, 10061661A 14035457 14035485 14046894 14035458 on . FDA classified it as Class II and gives the reason as "CareFusion is recalling the SmartSite Low Sorbing Infusion Set because they were assembled with a drip chamber that consist of a 15 micron filter. The Directions For Use did not indicate that the drip chamber has a 15...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2025-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- SmartSite Low Sorbing Infusion Set, Model Codes: 10014855A, 10015861A, 10061661A 14035457 14035485 14046894 14035458
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- CareFusion is recalling the SmartSite Low Sorbing Infusion Set because they were assembled with a drip chamber that consist of a 15 micron filter. The Directions For Use did not indicate that the drip chamber has a 15 micron filter and it does not provide the instruction that the set should not be used with blood administration due to the potential for occluding the 15 micron filter in the drip c
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- A recall letter is dated 6/25/14 and will be sent out to customers who purchased the SmartSite Low Sorbing Infusion Set because they were assembled with a drip chamber that consist of a 15 micron filter. The Directions For Use did not indicate that the drip chamber has a 15 micron filter and it does not provide the instruction that the set should not be used with blood administration due to the potential for occluding the 15 micron filter in the drip chamber. The letter informs the customers of the problems identified, potential risk, and actions to be taken. Customers are instructed to promptly complete and return the enclosed mandatory Customer Response Card to acknowledge the receipt of this communication and to expedite the corrective action process.
- Quantity in commerce
- 5,600 units
- Distribution
- Distributed in the states of FL, NM, WA, WI, and TX.
- Product codes and lots
- Lot No. 14035485 14046894 14035458 14035457
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2025-2014 | Class II | Terminated | Voluntary: Firm initiated |