Spacelabs Healthcare Ultraview SL Multigas Module, Model 92518 (Software V1.00.00-09). Provides a means of...
Spacelabs Healthcare Inc began a recall of Spacelabs Healthcare Ultraview SL Multigas Module, Model 92518 (Software V1.00.00-09). Provides a means of monitoring a variety of gas concentrations and alert... on . FDA classified it as Class II and gives the reason as "With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2077-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Spacelabs Healthcare Ultraview SL Multigas Module, Model 92518 (Software V1.00.00-09). Provides a means of monitoring a variety of gas concentrations and alert clinical personnel when the concentration of anesthetic agent(s), oxygen, carbon dioxide or nitrous oxide moves outside of user-defined limits.
- Product code
- CBR Analyzer, Gas, Nitrous-Oxide, Gaseous Phase (Anesthetic Conc.)
- Reason for recall
- With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Spacelabs sent an Urgent Medical Device Correction letter dated June 20, 20/14 to all affected customers. The letter identified the affected product, problem and the actions to be taken. Customers are informed that Spacelabs Field Service personnel will be contacting their facility to schedule a convenient time to confirm, at no cost, a software update is needed to resolve this issue. Customers with questions are instructed to contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
- Quantity in commerce
- 8 units total (3 units in US)
- Distribution
- Worldwide Distribution - US in the state of Georgia and in the countries of : Malaysia, Paraguay, and France.
- Product codes and lots
- Serial Numbers: 2518-000183 2518-000184 2518-000185 2518-000186 2518-000206 2518-000207 2518-000208 2518-000209
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2077-2014 | Class II | Terminated | Voluntary: Firm initiated |