Z-2118-2014Class II

Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the...

Medtronic Neuromodulation / initiated 2014-06-11 / Terminated / root cause: manufacturing process

Medtronic Neuromodulation began a recall of Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise... on . FDA classified it as Class II and gives the reason as "Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2118-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
Medtronic Nexdrive Micropositioning Drive. Models Ml-1000 and Ml-2000. For use in conjunction with the Medtronic Nexframe Stereotactic System for the precise positioning of microelectrodes and implantable leads. A stereotactic guidance system used in conjunction with Medtronic StealthStation Navigation Systems-image-guided surgery (IGS) systems-for Deep Brain Stimulation (DBS) procedures.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Potential for misalignment of the Z-stage scale. Using one of these devices for a procedure could result in the microelectrode being inserted to an incorrect target depth.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Medtronic sent an Urgent Medical Device Recall letter dated June 2014 on June 11, 2014, to all affected customers. The letter described the problem and the product involved in the recall. The letter described the "Action Requested" which included to stop using and return the recalled device and to complete and return the Reply Form. Customers with questions were instructed to contact Medtronic Neuromodulation Technical Services at 1-800-707-0933. EXPANDED 10/02/2014: Medtronic sent an Urgent Medical Device Recall letter beginning 10/02/2014. The letter explained that this letter is an update to the June 2014 action, to include 4 additional lot numbers. The letter explained the reason for recall identified all lots recalled and asked that devices stopped being used. A response form provided instructions on how to return product and the form was asked to be returned within 10 days. Questions should be directed to Medtronic Neuromodulation Technical Service at 1-800-707-0933, weekdays 7 am - 6 pm, Central time.
Quantity in commerce
300 EXPANDED 10/02/2014 130 devices
Distribution
Worldwide Distribution - USA including CA, CO, FL, IL, KS, KY, LA, MN, MT, NC, NE, OR, TX, VA, and UT, and Internationally to Canada, Spain, and United Kingdom.
Product codes and lots
Lot numbers: 082605614, 082601414, 082602014, 082634313, 082630913, 082629713, 082629613, 082626913, 082623113, 082621413, and 082618213.   EXPANDED 10/02/2014 Lot Numbers: 082616214, 082616314, 082618314, and 082618414.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2118-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2118-2014Class IITerminatedVoluntary: Firm initiated