Z-2169-2014Class II

5 mL Reaction Vessels (a consumable of the Abbott m2000sp Automated Fluid Handling System). The Abbott m2000...

Abbott Molecular, Inc. / initiated 2014-06-19 / Terminated / root cause: manufacturing process

Abbott Molecular began a recall of 5 mL Reaction Vessels (a consumable of the Abbott m2000sp Automated Fluid Handling System). The Abbott m2000 RealTime System is intended for use in performing... on . FDA classified it as Class II and gives the reason as "Abbott Molecular identified that some of the 5 ml Reaction Vessels (RVs) contained in lots 56944001, 57026001, 57059001, and 57673001 (original vendor lot 2198077) may have an insufficient rim. An RV with an...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2169-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Molecular, Inc.
Product
5 mL Reaction Vessels (a consumable of the Abbott m2000sp Automated Fluid Handling System). The Abbott m2000 RealTime System is intended for use in performing nucleic acid amplification testing by polymerase chain reaction in clinical laboratories. The system is comprised of the m2000rt and the m2000sp instruments. The Abbott m2000 RealTime System is intended for use in performing nucleic acid amplification testing by polymerase chain reaction in clinical laboratories. The system is comprised of the m2000rt and the m2000sp instruments.
Product code
JJH Clinical Sample Concentrator
Reason for recall
Abbott Molecular identified that some of the 5 ml Reaction Vessels (RVs) contained in lots 56944001, 57026001, 57059001, and 57673001 (original vendor lot 2198077) may have an insufficient rim. An RV with an insufficient rim may not be held securely in the 1 mL Subsystem Carrier. This may cause the RV to be seated incorrectly in the carrier or drop through it.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Abbott sent a Field Correction Recall / Urgent Field Safety Notice dated July 3, 2014, to all affected customers. June 19, 2014 were sent to all customers. The letters instructed customers to: 1) examine their inventories, 2) remove any of the affected lots, and 3) contact their Abbott Molecular Representativeat 1-800-553-7042 to coordinate replacement. Customers were also instructed to review this information with laboratory personnel and retain the communication for future reference. For questions regarding this recall call 224-361-7274
Quantity in commerce
1,688,000 reaction vessels
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to: AUSTRALIA AUSTRIA BELGIUM BENIN CAMEROON CHINA ETHIOPIA FINLAND FRANCE GERMANY INDIA INDONESIA IRELAND ISRAEL ITALY KENYA MALAYSIA MALAWI MALI NETHERLANDS NEW ZEALAND NORWAY PAKISTAN PHILIPPINES POLAND PORTUGAL RUSSIA SENEGAL SINGAPORE SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UKRAINE UNITED KINGDOM VIETNAM
Product codes and lots
Unique Product Identifier: 04J71-20;  Serial Numbers: 56944001, 57026001, 57059001, 57673001

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2169-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2169-2014Class IITerminatedVoluntary: Firm initiated