Z-2180-2014Class II

Siemens Syngo RT Therapist and Syngo RT Oncologist intended use of the SIEMENS branded Syngo RT Therapist and...

Siemens Medical Solutions USA, Inc. / initiated 2014-06-17 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens Syngo RT Therapist and Syngo RT Oncologist intended use of the SIEMENS branded Syngo RT Therapist and Syngo RT Oncologist software applications used... on . FDA classified it as Class II and gives the reason as "Siemens became aware of a potential issue with the syngo RT Therapist 4.3.SP1 and syngo RT Oncologist 4.3SP1 software applications running on Siemens brand LINAC systems that may experience sporadic system crashes that...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2180-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Syngo RT Therapist and Syngo RT Oncologist intended use of the SIEMENS branded Syngo RT Therapist and Syngo RT Oncologist software applications used with SIEMENS family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Siemens became aware of a potential issue with the syngo RT Therapist 4.3.SP1 and syngo RT Oncologist 4.3SP1 software applications running on Siemens brand LINAC systems that may experience sporadic system crashes that will cause the inability to unload patient data.
Root cause tag
software
FDA root cause
Software design
Action
In June, 2014, an urgent medical device correction field safety notice of improvement was sent to end users informing them of the product, problem, and action to be taken to avoid the potential safety issue.
Quantity in commerce
56
Distribution
UT, WI, NY
Product codes and lots
Syngo RT Therapist (RTT 4.3.SP1) - model no. 08162815 Syngo RT Oncologist (RTO 4.3.SP1) - model no. 10568604 Syngo RT Therapist (RTT 4.3.SP1) for U.S. affected units only:  10010 10046 10586  No Syngo RT Oncologist (RTO 4.3.SP1) listed for U.S.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2180-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2180-2014Class IITerminatedVoluntary: Firm initiated