Z-2037-2014Class II

GEMINI TF PET/CT16: 882470, GEMINI TF PET/CT 64: 882471, GEMINI TF Big Bore: 882476. A diagnostic imaging...

Philips North America LLC / initiated 2014-06-18 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of GEMINI TF PET/CT16: 882470, GEMINI TF PET/CT 64: 882471, GEMINI TF Big Bore: 882476. A diagnostic imaging system for fixed or mobile installations that... on . FDA classified it as Class II and gives the reason as "When beginning a procedure to start the gantry and CT Host, a Close Estop message may appear to the operator, but disappears before action can be taken. If this occurs it results in an inability to close E-Stop and the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2037-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
GEMINI TF PET/CT16: 882470, GEMINI TF PET/CT 64: 882471, GEMINI TF Big Bore: 882476. A diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
When beginning a procedure to start the gantry and CT Host, a Close Estop message may appear to the operator, but disappears before action can be taken. If this occurs it results in an inability to close E-Stop and the system is not operational for clinical use. This problem occurs randomly.
Root cause tag
software
FDA root cause
Software design
Action
Philips Healthcare is informing the consignee of this issue through a Field Safety Notice (FSN) 88200487-488. The US consignees will receive this notice via certified mail and Philips Key Market representatives will distribute the FSN internationally. Additionally, a Field Service Engineer will contact each site to schedule a time to install the software update.
Quantity in commerce
33
Distribution
Distributed in the states of CA, CO, IL, OK, OR and WI.
Product codes and lots
GEMINI TF PET/CT16: 882470 7019,7039,7099,7588,7593,7598,7599,7600,7604,7605,7596M  GEMINI TF PET/CT 64: 882471 7027,7044,7146,7153,7178,7182,7188,7578,7579,7585,7587,7589,7592,7594,7595,7602, 7603, 7606  GEMINI TF Big Bore: 882476 9019,9227,9232,9233

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2037-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2037-2014Class IITerminatedVoluntary: Firm initiated